Department of Radiation Oncology, Shanghai Chest Hospital, Shanghai Jiao Tong University, Shanghai, China.
Shcool of Physics and Technology, University of Wuhan, Wuhan, China.
J Appl Clin Med Phys. 2020 Dec;21(12):74-83. doi: 10.1002/acm2.13061. Epub 2020 Oct 31.
Position accuracy of the multi-leaf collimator (MLC) is essential in stereotactic body radiotherapy (SBRT). This study is aimed to investigate the dosimetric impacts of the MU-weighted MLC positioning uncertainties of SBRT for patients with early stage peripheral non-small cell lung cancer (NSCLC).
Three types of MLC position error were simulated: Type 1, random error; Type 2, system shift, in which both MLC banks shifted to the left or right direction; and Type 3, in which both MLC banks moved with same magnitudes in the opposite directions. Two baseline plans were generated: an automatic plan (AP) and a manually optimized plan (MP). Multi-leaf collimator position errors were introduced to generate simulated plans with the preset MLC leaf position errors, which were then reimported into the Pinnacle system to generate simulated plans, respectively. The dosimetric parameters (CI, nCI, GI, etc.) and gEUD values of PTV and OARs were calculated. Linear regression between MU-weighted/unweighted MLC position error and gEUD was performed to obtain dose sensitivity.
The dose sensitivities of the PTVs were -4.93, -38.94, -41.70, -55.55, and 30.33 Gy/mm for random, left shift, right shift, system close, and system open MLC errors, respectively. There were significant differences between the MU-weighted and the unweighted dose sensitivity, which was -38.94 Gy/mm vs -3.42 Gy/mm (left shift), -41.70 Gy/mm vs -3.56 Gy/mm (right shift), -55.55 Gy/mm vs -4.84 Gy/mm (system close), and 30.33 vs 2.64 Gy/mm (system open). For the system open/close MLC errors, as the PTV volume became larger, the dose sensitivity decreased. APs provided smaller dose sensitivity for the system shift and system close MLC errors compared to the conventional MPs.
There was significant difference in dose sensitivity between MU-weighted and unweighted MLC position error of SBRT radiotherapy in peripheral NSCLC. MU is suggested to be included in the dosimetric evaluation of the MLC misalignments, since it is much closer to clinical radiotherapy.
多叶准直器(MLC)的位置精度在立体定向体放射治疗(SBRT)中至关重要。本研究旨在探讨早期外周非小细胞肺癌(NSCLC)患者 SBRT 中 MLC 定位不确定性的 MU 加权对剂量的影响。
模拟了三种 MLC 位置误差类型:类型 1,随机误差;类型 2,系统移位,其中两个 MLC 叶片向左或向右移动;类型 3,两个 MLC 叶片以相同的幅度向相反的方向移动。生成了两个基线计划:自动计划(AP)和手动优化计划(MP)。通过预设的 MLC 叶片位置误差引入多叶准直器位置误差,分别生成模拟计划,并将其重新导入 Pinnacle 系统生成模拟计划。计算了 PTV 和 OAR 的剂量学参数(CI、nCI、GI 等)和 gEUD 值。对 MU 加权/未加权 MLC 位置误差与 gEUD 进行线性回归,以获得剂量敏感性。
PTV 的剂量敏感性分别为随机、左移、右移、系统关闭和系统打开 MLC 误差的-4.93、-38.94、-41.70、-55.55 和 30.33 Gy/mm。MU 加权与未加权剂量敏感性之间存在显著差异,分别为-38.94 Gy/mm 与-3.42 Gy/mm(左移)、-41.70 Gy/mm 与-3.56 Gy/mm(右移)、-55.55 Gy/mm 与-4.84 Gy/mm(系统关闭)和 30.33 Gy/mm 与 2.64 Gy/mm(系统打开)。对于系统打开/关闭 MLC 误差,随着 PTV 体积的增大,剂量敏感性降低。AP 对系统移位和系统关闭 MLC 误差的剂量敏感性比对常规 MPs 小。
在外周 NSCLC 的 SBRT 放疗中,MU 加权与未加权 MLC 位置误差的剂量敏感性存在显著差异。由于它更接近临床放射治疗,因此建议在 MLC 对准误差的剂量评估中包含 MU。