Srinivasan V Ramesh, Rekha M, Edsor Effie, Raja S Pradheep, Kumar T Dinesh, Kalaiselvan S
Department of Oral and Maxillofacial Surgery, RVS Dental College and Hospital, Coimbatore, Tamil Nadu, India.
Department of Oral and Maxillofacial Pathology, RVS Dental College and Hospital, Coimbatore, Tamil Nadu, India.
J Pharm Bioallied Sci. 2020 Aug;12(Suppl 1):S607-S612. doi: 10.4103/jpbs.JPBS_163_20. Epub 2020 Aug 28.
The aim of this study was to assess the quality of platelet-rich plasma (PRP) produced by in-house desktop centrifuge method and compare it with that of standardized commercial PRP.
REMI desktop centrifuge was used to prepare PRP and to compare with standardized commercial PRP by calculating the quantity of platelets using Beckman Coulter cell counter in 10 PRP samples and assessing the morphological quality of platelets using JEOL JEM transmission electron microscope (TEM).
The test for platelet count in desktop PRP with the test value of therapeutic PRP was 12.618. The value was <0.001, which was significant statistically. The data followed a normal distribution in normal Q-Q plot for platelet count in desktop centrifuge. So the test samples were not much deviated.
The platelet count was lesser than that of standardized commercial PRP. When viewed under JEOL JEM transmission electron microscope, the α granules in platelets were intact and the morphological quality of the PRP was good.
With this study, we have determined that the morphological quality of PRP produced by the in-house desktop centrifuge method is comparable to that of standardized commercial PRP. Though the quantity of platelets was less than 1 million cells/μL, the clinical results were good with desired bone formation, thereby providing good avenue for further research.
本研究旨在评估通过内部台式离心机方法制备的富血小板血浆(PRP)的质量,并将其与标准化商业PRP的质量进行比较。
使用雷米台式离心机制备PRP,并通过使用贝克曼库尔特细胞计数器计算10个PRP样本中的血小板数量,以及使用日本电子JEM透射电子显微镜(TEM)评估血小板的形态质量,与标准化商业PRP进行比较。
台式PRP中血小板计数与治疗性PRP测试值的检验值为12.618。P值<0.001,具有统计学意义。在台式离心机血小板计数的正态Q-Q图中,数据呈正态分布。因此,测试样本偏差不大。
血小板计数低于标准化商业PRP。在日本电子JEM透射电子显微镜下观察,血小板中的α颗粒完整,PRP的形态质量良好。
通过本研究,我们确定通过内部台式离心机方法制备的PRP的形态质量与标准化商业PRP相当。尽管血小板数量少于100万个细胞/μL,但临床结果良好,骨形成理想,从而为进一步研究提供了良好的途径。