Suppr超能文献

带双吸气环管的流量同步 NIPPV:一项体外研究。

Flow-synchronized NIPPV with double-inspiratory loop cannula: An in vitro study.

机构信息

Emeritus Consultant in Paediatrics, Policlinico Umberto I, Sapienza University of Rome, Rome, Italy.

Departmet of Pediatrics, Neonatal Intensive Care Unit and Neonatology, "V. Buzzi"-Ospedale dei Bambini-ASST-FBF-Sacco, Milan, Italy.

出版信息

Pediatr Pulmonol. 2021 Feb;56(2):400-408. doi: 10.1002/ppul.25161. Epub 2020 Nov 19.

Abstract

BACKGROUND

Although short binasal prongs (SBP) are the most common interface for noninvasive ventilation, the "double-inspiratory loop cannulas" (DILC) have recently been introduced into many neonatal intensive care units. DILC show advantages over SBP, including reduced nasal trauma and increased comfort. However, their higher intrinsic resistance may compromise ventilation. Our aim was to test a new, low resistance DILC interface.

METHODS

A test lung was programmed to simulate preterm neonates (500-2000 g BW) with moderate-to-severe respiratory distress syndrome. The artificial nose was designed to keep prongs-to-nares leaks to around 30%. Giulia® ventilator (GINEVRI srl) was used to provide nasal continuous positive airway pressure (NCPAP) and flow synchronized nasal intermittent positive pressure ventilation (NIPPV). NCPAP was set at 4-10 cmH O and synchronized-NIPPV (SNIPPV) at peak inspiratory pressure, 15-20-25 cmH O; inspiratory time, 0.3-0.5 s; and positive end-expiratory pressure, 5-8 cmH O. Four sizes of Sync-flow Cannula® (GINEVRI srl) were tested. The Sync-flow Cannula® was compared with Neotech RAM® cannula and Ginevri SBP®. The outcome measures were the flow/pressure relationship through the four Sync-flow Cannula® sizes, the difference in resistance, the drop in ventilator-alveoli pressure measured by the test lung and the system response time during flow-SNIPPV.

RESULTS

The smaller DILC sizes had the lowest flow-pressure ratio. The resistance of the RAM® cannula was significantly higher compared to the other interfaces (p < .001). With 30% leaks, there was a 4-38% ventilator-alveoli drop in pressure, depending on interface size. The system response time was excellent (~65-70 ms).

CONCLUSIONS

With about 30% leaks, the Sync-flow Cannula® interfaces result in good pressure transmission and give optimal performance for flow-SNIPPV. Clinical studies are needed to confirm the clinical relevance of these data.

摘要

背景

虽然短鼻夹(SBP)是无创通气最常用的接口,但“双吸气环管”(DILC)最近已被引入许多新生儿重症监护病房。DILC 与 SBP 相比具有优势,包括减少鼻损伤和增加舒适度。然而,它们更高的固有阻力可能会影响通气。我们的目的是测试一种新的、低阻力 DILC 接口。

方法

测试肺被编程模拟具有中重度呼吸窘迫综合征的早产儿(500-2000gBW)。人工鼻设计将鼻夹到鼻腔的泄漏保持在 30%左右。使用 Giulia®呼吸机(GINEVRI srl)提供经鼻持续气道正压通气(NCPAP)和流量同步经鼻间歇正压通气(NIPPV)。NCPAP 设置为 4-10cmH2O,同步-NIPPV(SNIPPV)设置为吸气峰压 15-20-25cmH2O;吸气时间 0.3-0.5s;呼气末正压 5-8cmH2O。测试了四种尺寸的 Sync-flow Cannula®(GINEVRI srl)。将 Sync-flow Cannula®与 Neotech RAM®管和 Ginevri SBP®进行了比较。测量指标为四种 Sync-flow Cannula®尺寸的流量/压力关系、阻力差异、测试肺和系统响应时间测量的呼吸机肺泡压力下降以及在流量-SNIPPV 期间的系统响应时间。

结果

较小的 DILC 尺寸具有最低的流量/压力比。与其他接口相比,RAM®管的阻力明显更高(p<.001)。在有 30%泄漏的情况下,根据接口尺寸,呼吸机肺泡压力下降 4-38%。系统响应时间非常出色(~65-70ms)。

结论

在有 30%泄漏的情况下,Sync-flow Cannula®接口可实现良好的压力传递,并为流量-SNIPPV 提供最佳性能。需要进行临床研究以确认这些数据的临床相关性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验