Braillon P, Duboeuf F, Delmas P D, Meunier P J
Département de Physique, Université Claude Bernard, Villeur-banne, France.
Pathol Biol (Paris). 1987 Nov;35(9):1205-13.
Measurements were made with a DPA apparatus (Novo Lab 22a) on different phantoms and on volunteers in an attempt to evaluate the system precision. The reproductibility was found in the range of 0.98 to 4.10% in the case of in vitro measurements, depending on the geometry of the phantoms used, and in the range of 1.6 to 2.94% for volunteers after repositioning. Secondly, the BMD in the lumbar spine of normal women and normal men was estimated. In control females, the BMD is well fitted to the age by a cubic regression. The maximum value of the BMD is found in this case at the age of 31.5 and the maximum rate of bone loss takes place at 57. Total bone loss between 31.5 and the elderly is about 32%. In control males, results are more scattered and are represented by a simple linear regression. The average mineral loss between 30 and 80 years is 11.5% in this area of measurement.
使用DPA仪器(Novo Lab 22a)对不同的体模和志愿者进行测量,以评估该系统的精度。体外测量时,根据所用体模的几何形状,再现性在0.98%至4.10%的范围内;志愿者重新定位后,再现性在1.6%至2.94%的范围内。其次,估计了正常女性和正常男性腰椎的骨密度。在对照女性中,骨密度通过三次回归很好地拟合年龄。在这种情况下,骨密度的最大值出现在31.5岁,最大骨丢失率发生在57岁。31.5岁至老年之间的总骨丢失约为32%。在对照男性中,结果更为分散,由简单线性回归表示。在该测量区域,30至80岁之间的平均矿物质丢失为11.5%。