Reference values for trabecular and cortical vertebral bone density in single and dual-energy quantitative computed tomography.

作者信息

Kalender W A, Felsenberg D, Louis O, Lopez P, Klotz E, Osteaux M, Fraga J

机构信息

Siemens Medical Systems, Erlangen, Germany.

出版信息

Eur J Radiol. 1989 May;9(2):75-80.

PMID:2743986
Abstract

The validity of reference values for vertebral bone mineral density (BMD) reported in the literature has frequently been questioned. The introduction of dual-energy methods, new calibration systems, and automated image evaluation algorithms, has also made it necessary for us to re-evaluate BMD values in a normal population. We present patient selection criteria, CT measurement and evaluation protocols and results of single energy (SEQCT) and dual energy quantitative CT (DEQCT) for trabecular (tBMD) and cortical (cBMD) bone density in 135 males and 139 females. For SEQCT, tBMD loss rates of 1.7 and 1.8 mg/ml per year for males and females, respectively, were determined by linear regression. The respective DEQCT data show smaller decrements with age. Different statistical models for tBMD in females were tested; models based on a multiphase variation of tBMD yielded an improvement of the fit. For cBMD the percent rate of loss with age is much smaller than for tBMD; females showed a significantly higher loss rate than males, however. A comparison of our results with published data is included.

摘要

文献中报道的脊椎骨矿物质密度(BMD)参考值的有效性经常受到质疑。双能方法、新的校准系统和自动图像评估算法的引入,也使我们有必要重新评估正常人群的BMD值。我们介绍了患者选择标准、CT测量和评估方案,以及135名男性和139名女性的小梁骨(tBMD)和皮质骨(cBMD)密度的单能(SEQCT)和双能定量CT(DEQCT)结果。对于SEQCT,通过线性回归确定男性和女性每年的tBMD丢失率分别为1.7和1.8mg/ml。相应的DEQCT数据显示,随着年龄增长,下降幅度较小。对女性tBMD的不同统计模型进行了测试;基于tBMD多阶段变化的模型拟合度有所提高。对于cBMD,随年龄增长的丢失率百分比远低于tBMD;然而,女性的丢失率明显高于男性。本文还将我们的结果与已发表的数据进行了比较。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验