Kim Hyeonwoo, Joo Yoon Ha, Yu Na Hee, Kwon Sung Tack, Lee Jung Chan, Kim Byung Jun
Department of Plastic and Reconstructive Surgery, Seoul National University Hospital.
Department of Biomedical Engineering, Seoul National University College of Medicine.
Medicine (Baltimore). 2020 Nov 20;99(47):e23149. doi: 10.1097/MD.0000000000023149.
Digital tourniquets are widely used for the management of digital injuries in emergency departments or outpatient clinics. This study is focused on the pressure analysis of digital tourniquets on some points not covered in the existing literature.A total of thirty volunteers were enrolled in this study. Instantaneous surface pressure was measured at the thumbs, index fingers, and little fingers. We investigated the pressure according to the circumference of digits, tourniquet types, and measurement sites (dorsal and mid lateral volar sides) above the digital vessels. Continuous pressure was measured in artificial silicone models to determine the change of pressure over 2 hours.The average pressure measured on the mid lateral volar side of volunteers fingers were 154.3 ± 54.9 mm Hg by T-Ring and 162.6 ± 61.0 mm Hg by surgical glove. The pressure on the dorsal side were 224.7 ± 57.7 mm Hg by T-Ring and 228.8 ± 66.0 mm Hg by surgical glove, each significantly higher than the mid lateral volar side. The circumference of digits did not significantly affect the surface pressure. The pressure pattern did not change significantly over 2 hours in both tourniquet types.The surface pressure of the mid lateral volar side was significantly lower than that of the dorsal side. However, there was no significant pressure difference according to the circumference of digits. Time dependent pressure change were not significantly different between 2 tourniquets.
数字止血带在急诊科或门诊中广泛用于处理手指损伤。本研究聚焦于数字止血带在现有文献未涉及的某些方面的压力分析。共有30名志愿者参与本研究。在拇指、食指和小指处测量瞬时表面压力。我们根据手指周长、止血带类型以及手指血管上方的测量部位(掌侧背侧和中外侧)来研究压力。在人工硅胶模型中测量持续压力,以确定2小时内压力的变化。志愿者手指中外侧掌侧通过T形环测量的平均压力为154.3±54.9毫米汞柱,通过手术手套测量为162.6±61.0毫米汞柱。背侧通过T形环测量的压力为224.7±57.7毫米汞柱,通过手术手套测量为228.8±66.0毫米汞柱,均显著高于中外侧掌侧。手指周长对表面压力无显著影响。两种止血带类型在2小时内压力模式均无显著变化。中外侧掌侧的表面压力显著低于背侧。然而,根据手指周长,压力无显著差异。两种止血带之间随时间的压力变化无显著差异。