数字止血带:压力与疼痛感知的比较分析

Digital tourniquets: a comparative analysis of pressures and pain perception.

作者信息

Naim Soulat, Srinivasan Makaram S

机构信息

Department of Orthopaedics, Blackburn Royal Infirmary, Blackburn, UK.

出版信息

Acta Orthop Belg. 2008 Apr;74(2):195-9.

Abstract

Digital tourniquets have been condemned because of the reported occurrence of neurovascular complications. This study compares the pressures and pain perception between three different tourniquets, namely the rubber glove finger stall, Foley's urinary catheter and the commercial band tourniquet. The subjects involved were 20 volunteers with 80 fingers studied in total. A direct pressure measuring technique was used. The pressures recorded were highest and most variable with the catheter tourniquet, whereas the rubber glove tourniquet recorded the lowest pressures. Correspondingly, visual analogue scale in relation to patient discomfort showed high scores with the catheter tourniquet and low scores with the rubber glove tourniquet. We conclude that catheter tourniquet use should be avoided as it generates extreme and variable pressures, whereas rubber glove finger stall tourniquets appear as a better alternative.

摘要

由于有报道称使用数字止血带会出现神经血管并发症,因此其已受到谴责。本研究比较了三种不同止血带(即橡胶手套指套、Foley导尿管和商用束带止血带)之间的压力和疼痛感知。研究对象为20名志愿者,共研究了80根手指。采用了直接压力测量技术。记录到的压力在导尿管止血带使用时最高且变化最大,而橡胶手套止血带记录的压力最低。相应地,与患者不适相关的视觉模拟量表显示,导尿管止血带得分高,橡胶手套止血带得分低。我们得出结论,应避免使用导尿管止血带,因为它会产生极高且变化不定的压力,而橡胶手套指套止血带似乎是更好的选择。