Institut de physique du globe de Paris, CNRS, Université de Paris, 75005, Paris, France.
Laboratoire Cogitamus, 1 ¾ rue Descartes, 75005, Paris, France.
Anal Bioanal Chem. 2021 Jan;413(3):923-933. doi: 10.1007/s00216-020-03048-y. Epub 2020 Nov 24.
Measurement of nanoparticle (NP) concentration and size by single-particle inductively coupled plasma mass spectrometry (spICP-MS) usually requires the use of a NP reference material to determine the loss of NPs and/or ions during their transport from the sample solution to the detection system. The determination of this loss, qualified as nebulization efficiency (η) and/or transport efficiency (η), is time-consuming, costly and lacks reliability. Nebulization of the NPs directly into the plasma (without a spray chamber) results in η = 100% and is thus a promising strategy to avoid these calibration steps. In this work, we used the μ-dDIHEN introduction system: a demountable direct injection high-efficiency nebulizer (dDIHEN) hyphenated to a flow-injection valve and a gas displacement pump. For the first time with a continuous flow nebulizer, complete transport efficiency was reached (i.e. η = 100%). Operated at a very low uptake rate (as low as 8 μL min), the μ-dDIHEN accurately and reproducibly determined average diameters of Au-, Ag- and Pt-NPs, in full agreement with their reference values. It was also successfully tested for Au-NPs in complex matrices, such as surface waters. spICP-MS analyses with the μ-dDIHEN sample introduction system only require a dissolved standard calibration to determine NP average diameter (d in nm) and number concentration (N) from the simplified set of equations: [Formula: see text] and [Formula: see text]Graphical abstract.
采用单颗粒电感耦合等离子体质谱(spICP-MS)测量纳米颗粒(NP)的浓度和粒径,通常需要使用 NP 参考物质来确定 NP 和/或离子在从样品溶液传输到检测系统的过程中的损失。这种损失的确定,被称为雾化效率(η)和/或传输效率(η),既耗时又昂贵,而且缺乏可靠性。将 NPs 直接雾化到等离子体中(不使用喷雾室)可实现η=100%,因此这是一种避免这些校准步骤的有前途的策略。在这项工作中,我们使用了 μ-dDIHEN 引入系统:一种可拆卸的直接注射高效雾化器(dDIHEN)与流动注射阀和气体置换泵联用。这是首次在连续流动雾化器上实现了完全的传输效率(即 η=100%)。在非常低的吸收速率(低至 8 μL min)下,μ-dDIHEN 可以准确且可重现地测定 Au、Ag 和 Pt-NP 的平均粒径,与它们的参考值完全一致。它还成功地测试了复杂基质(如地表水)中的 Au-NP。采用 μ-dDIHEN 样品引入系统的 spICP-MS 分析仅需要溶解标准校准,即可从简化的方程组中确定 NP 平均直径(d in nm)和数浓度(N):[公式:见文本]和[公式:见文本]。