全身正电子发射断层扫描/计算机断层扫描中计算机断层扫描部分的辐射剂量调制不佳。

SUBOPTIMAL MODULATION OF RADIATION DOSE IN THE COMPUTED TOMOGRAPHY COMPONENT OF WHOLE-BODY POSITRON EMISSION TOMOGRAPHY/COMPUTED TOMOGRAPHY.

机构信息

Department of Diagnostic Radiology, Kitasato University School of Medicine, 1-15-1 Kitasato, Minami-ku, Sagamihara, Kanagawa, Japan.

Department of Radiology, Kitasato University Hospital, 1-15-1 Kitasato, Minami-ku, Sagamihara, Kanagawa, Japan.

出版信息

Radiat Prot Dosimetry. 2020 Dec 30;192(1):69-74. doi: 10.1093/rpd/ncaa197.

Abstract

Radiation exposure in computed tomography (CT) is automatically modulated by automatic exposure control (AEC) mainly based on scout images. To simulate the whole-body positron emission tomography/CT, CT images of a phantom were obtained using the posteroanterior scout image alone (PA scout) or the posteroanterior and lateral images (PA + Lat scout). Old and new versions of the AEC software were compared. Using the old version of the software and the PA scout, a markedly high dose at the top of the head was observed, which varied depending on the position of the phantom. This issue was resolved in the new version of the software. Radiation dose in the shoulder region was much higher using the PA scout than using the PA + Lat scout, even with the new version of the software. AEC may cause unreasonably high radiation exposure locally, and the appropriateness of the dose modulation pattern should be examined at each facility.

摘要

计算机断层扫描(CT)中的辐射暴露主要通过自动曝光控制(AEC)基于定位像自动调节。为了模拟全身正电子发射断层扫描/CT,仅使用前后位定位像(PA 定位像)或前后位和侧位定位像(PA+Lat 定位像)获取了体模的 CT 图像。对新旧版本的 AEC 软件进行了比较。使用旧版本软件和 PA 定位像,观察到头部顶部的剂量明显升高,这取决于体模的位置。在新版本软件中解决了这一问题。使用 PA 定位像时,肩部区域的辐射剂量明显高于使用 PA+Lat 定位像,即使使用新版本软件也是如此。AEC 可能会导致局部不合理的高辐射暴露,应在每个医疗机构检查剂量调制模式的适宜性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验