Hanks G A, Foster W C, Cardea J A
Department of Surgery, Medical College of Virginia, Virginia Commonwealth University, Richmond 23298-0001.
Clin Orthop Relat Res. 1988 Jan(226):206-18.
Fifty femur fractures were treated with the Brooker-Wills intramedullary locking nail. The indications included subtrochanteric, subisthmal, segmental or comminuted acute fractures or nonunions, and intramedullary shortening procedures. Eighteen technical problems in 13 (26%) patients were encountered during insertion of the nail. The incidence of technical problems was high early in the series and in procedures done with the patient in the supine position. Adverse clinical results from technical errors occurred in two patients. Most technical errors seem avoidable with careful technique. The mean healing time was only 12.1 weeks and there were four nonunions (8%). One patient had a 10 degrees malunion, four patients had 1-2 cm of shortening, and there were two late device fractures. No patients had rotational malalignment or deep infections. The results and complication rate are comparable to those reported for other interlocking systems. The Brooker-Wills nail is useful for treatment of complex femur fractures and has been effective in preventing malrotation, angulation, and excessive shortening.
采用布鲁克-威尔斯髓内锁定钉治疗了50例股骨骨折。适应证包括转子下、峡部以下、节段性或粉碎性急性骨折或骨不连,以及髓内缩短手术。在插入髓内钉过程中,13例(26%)患者出现了18个技术问题。技术问题的发生率在该系列手术早期以及患者仰卧位进行的手术中较高。两名患者因技术失误出现了不良临床结果。大多数技术失误通过仔细操作似乎是可以避免的。平均愈合时间仅为12.1周,有4例骨不连(8%)。1例患者出现10度畸形愈合,4例患者缩短1 - 2厘米,有2例出现后期内固定断裂。没有患者出现旋转畸形或深部感染。结果和并发症发生率与其他交锁系统报道的相当。布鲁克-威尔斯钉对于治疗复杂股骨骨折有用,并且在预防旋转畸形、成角畸形和过度缩短方面一直有效。