即时检测粪便血红蛋白免疫胶体金法的分析性能评估。

Assessment of the analytical performance of point-of-care faecal immunochemical tests for haemoglobin.

机构信息

Bowel Cancer Screening Programme, Southern Hub, Royal Surrey County Hospital, Guildford, UK.

Berkshire and Surrey Pathology Services, Guildford, UK.

出版信息

Ann Clin Biochem. 2021 May;58(3):181-189. doi: 10.1177/0004563220986595. Epub 2021 Jan 12.

Abstract

BACKGROUND

The faecal immunochemical test (FIT) detects the presence of haemoglobin (Hb) in faeces. It is used as a screening tool for colorectal cancer (CRC) and increasingly to triage patients presenting with symptoms of CRC. A number of quantitative point-of-care (POC) FIT systems marketed for professional use and intended for use in a clinical setting are available. Here we reviewed the POC FIT systems available; three (Eurolyser Cube, OC-Sensor iO and QuikRead go) were evaluated to assess their performance against manufacturers' claims and suitability for use in a clinical setting.

METHODS

The analytical evaluation of the POC FIT systems was undertaken using Hb lysates, patient samples and an external quality assessment sample. The evaluation focused on linearity, recovery, imprecision, prozone effect, Hb variant detection and suitability for use in a clinical setting.

RESULTS

All three POC FIT systems performed to their manufacturer's claims and demonstrated good analytical performance with acceptable linearity, recovery, within- and between-run imprecision. The QuikRead go and OC-Sensor iO were able to accurately detect samples with results above their measuring range. However, because of a prozone effect the Eurolyser Cube gave falsely low results when using high concentrations of Hb. The QuikRead go performed best in the usability assessment due to portability and timeliness of result.

CONCLUSION

Each system performed according to their manufacturers' claims. The QuikRead go and OC-Sensor iO are suitable for use. The Eurolyser Cube is not recommended because of the risk of falsely low results.

摘要

背景

粪便免疫化学检测(FIT)检测粪便中血红蛋白(Hb)的存在。它被用作结直肠癌(CRC)的筛查工具,并且越来越多地用于对出现 CRC 症状的患者进行分诊。有许多商业化的定量即时检测(POC)FIT 系统,可用于专业用途,并旨在在临床环境中使用。在这里,我们对现有的 POC FIT 系统进行了综述;评估了三种(Eurolyser Cube、OC-Sensor iO 和 QuikRead go),以评估它们与制造商声称的性能以及在临床环境中的适用性。

方法

使用 Hb 裂解物、患者样本和外部质量评估样本对 POC FIT 系统进行了分析评估。评估重点是线性、回收率、精密度、前区效应、Hb 变体检测和在临床环境中的适用性。

结果

所有三种 POC FIT 系统均符合制造商的声称,并表现出良好的分析性能,具有可接受的线性、回收率、批内和批间精密度。QuikRead go 和 OC-Sensor iO 能够准确检测结果高于其测量范围的样本。然而,由于前区效应,Eurolyser Cube 在使用高浓度 Hb 时会给出错误的低结果。由于便携性和及时获得结果,QuikRead go 在可用性评估中表现最佳。

结论

每个系统都符合其制造商的声称。QuikRead go 和 OC-Sensor iO 适合使用。由于存在错误低值的风险,不建议使用 Eurolyser Cube。