Suppr超能文献

一种基于电离测量的驻极体被动式环境氡-222监测仪。

An electret passive environmental 222Rn monitor based on ionization measurement.

作者信息

Kotrappa P, Dempsey J C, Hickey J R, Stieff L R

机构信息

Plastic Casting Inc., Olean, NY 14760.

出版信息

Health Phys. 1988 Jan;54(1):47-56. doi: 10.1097/00004032-198801000-00002.

Abstract

The electret passive environmental 222Rn monitor (E-PERM) is an extension of electret dosimeters used for measurement of x and gamma radiation. An E-PERM consists of a small cup or canister, having an electret at the bottom, and a filtered inlet at the top. The 222Rn gas entering through the filter and the decay products formed inside the cup generate ions which are collected by the electret. The reduction of charge (or surface potential) on the electret is a measure of time integrated 222Rn exposure. An E-PERM of 220-mL volume with an electret of 0.23 cm thickness gave a surface potential drop of 2.5 V for 37 Bq m-3 d (1 pCi L-1 d). The electret voltage was measured with a specially built surface potential voltmeter. This sensitivity was found adequate for a 1-wk measurement of 222Rn in homes. For longer term measurements, an E-PERM of 40-mL volume and an electret of 51-micron thickness was developed which gave a surface potential drop of 2.6 V for 37 Bq m-3 y (1 pCi L-1 y). Other combinations of chamber volume and electret thicknesses gave responses between these two values. The surface potential of electrets made from Teflon FEP were shown to stay stable even under extreme conditions of relative humidity. The ion collection process in E-PERMs was also shown to be independent of humidity down to an electret surface potential of 100 V.

摘要

驻极体被动式环境222Rn监测仪(E-PERM)是用于测量X射线和γ射线的驻极体剂量计的扩展。一个E-PERM由一个小杯子或罐组成,底部有一个驻极体,顶部有一个过滤入口。通过过滤器进入的222Rn气体以及在杯子内部形成的衰变产物会产生离子,这些离子被驻极体收集。驻极体上电荷(或表面电位)的减少是对时间积分的222Rn暴露的一种度量。一个体积为220 mL、驻极体厚度为0.23 cm的E-PERM,对于37 Bq m-3 d(1 pCi L-1 d)的情况,表面电位下降了2.5 V。驻极体电压用专门制造的表面电位电压表进行测量。发现这种灵敏度足以在家中进行为期1周的222Rn测量。对于长期测量,开发了一种体积为40 mL、驻极体厚度为51微米的E-PERM,对于37 Bq m-3 y(1 pCi L-1 y)的情况,表面电位下降了2.6 V。腔室体积和驻极体厚度的其他组合给出了介于这两个值之间的响应。由聚四氟乙烯FEP制成的驻极体的表面电位即使在极端相对湿度条件下也能保持稳定。E-PERMs中的离子收集过程也被证明在驻极体表面电位低至100 V时与湿度无关。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验