Diurnal variation of corneal elasticity in healthy young human using air-puff optical coherence elastography.

机构信息

School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, China.

Department of Biomedical Engineering, College of Engineering, Peking University, Wenzhou, Beijing, China.

出版信息

J Biophotonics. 2021 Aug;14(8):e202000440. doi: 10.1002/jbio.202000440. Epub 2021 May 16.

Abstract

Due to the disruption of intraocular pressure (IOP) and central corneal thickness (CCT), diurnal variation in normal young human corneal elasticity is not clear. Using the custom-built air-puff optical coherence elastography, one eye of 21 normal subjects is enrolled randomly to measure the central corneal elasticity, IOP, and CCT in different time points within a day. Based on the multi-level model, the corneal elastic modulus is found to have a linear positive relation with IOP (P < .01) but not CCT (P = .175) and time point (P = .174-.686). A new indicator, corneal elasticity change rate, is proposed to present the magnitude of corneal elasticity change caused by 1 mmHg IOP, which can correct the interference effect of IOP. The results show that the corneal elasticity in the normal young human does not have the characteristics of diurnal variation under IOP control. Furthermore, IOP plays an important role in the corneal elasticity, and corneal elasticity change rate can increase the comparability of results between individuals.

摘要

由于眼内压(IOP)和中央角膜厚度(CCT)的干扰,正常年轻人类角膜弹性的昼夜变化尚不清楚。本研究使用定制的空气脉冲光学相干弹性成像技术,随机招募 21 名正常受试者的一只眼睛,在一天内的不同时间点测量中央角膜弹性、IOP 和 CCT。基于多级模型,角膜弹性模量与 IOP 呈线性正相关(P<.01),但与 CCT(P=.175)和时间点(P=.174-.686)无关。本研究提出了一个新的指标——角膜弹性变化率,用于表示 1mmHg IOP 引起的角膜弹性变化幅度,可以校正 IOP 的干扰效应。结果表明,在 IOP 控制下,正常年轻人的角膜弹性没有昼夜变化的特征。此外,IOP 在角膜弹性中起重要作用,而角膜弹性变化率可以提高个体之间结果的可比性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验