五维定量低剂量多任务动态对比增强 MRI:乳腺癌的初步研究。

Five-dimensional quantitative low-dose Multitasking dynamic contrast- enhanced MRI: Preliminary study on breast cancer.

机构信息

Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA.

Department of Bioengineering, University of California, Los Angeles, California, USA.

出版信息

Magn Reson Med. 2021 Jun;85(6):3096-3111. doi: 10.1002/mrm.28633. Epub 2021 Jan 11.

Abstract

PURPOSE

To develop a low-dose Multitasking DCE technique (LD-MT-DCE) for breast imaging, enabling dynamic T mapping-based quantitative characterization of tumor blood flow and vascular properties with whole-breast coverage, a spatial resolution of 0.9 × 0.9 × 1.1 mm , and a temporal resolution of 1.4 seconds using a 20% gadolinium dose (0.02 mmol/kg).

METHODS

Magnetic resonance Multitasking was used to reconstruct 5D images with three spatial dimensions, one T recovery dimension for dynamic T quantification, and one DCE dimension for contrast kinetics. Kinetic parameters , , , and were estimated from dynamic T maps using the two-compartment exchange model. The LD-MT-DCE repeatability and agreement against standard-dose MT-DCE were evaluated in 20 healthy subjects. In 7 patients with triple-negative breast cancer, LD-MT-DCE image quality and diagnostic results were compared with that of standard-dose clinical DCE in the same imaging session. One-way unbalanced analysis of variance with Tukey test was performed to evaluate the statistical significance of the kinetic parameters between control and patient groups.

RESULTS

The LD-MT-DCE technique was repeatable, agreed with standard-dose MT-DCE, and showed excellent image quality. The diagnosis using LD-MT-DCE matched well with clinical results. The values of , , and were significantly different between malignant tumors and normal breast tissue (P < .001, < .001, and < .001, respectively), and between malignant and benign tumors (P = .020, .003, and < .001, respectively).

CONCLUSION

The LD-MT-DCE technique was repeatable and showed excellent image quality and equivalent diagnosis compared with standard-dose clinical DCE. The estimated kinetic parameters were capable of differentiating between normal breast tissue and benign and malignant tumors.

摘要

目的

开发一种用于乳腺成像的低剂量多任务动态对比增强技术(LD-MT-DCE),实现基于动态 T 映射的全乳覆盖、空间分辨率为 0.9×0.9×1.1mm 和时间分辨率为 1.4 秒的肿瘤血流和血管特性的定量特征描述,所用钆剂量为 20%(0.02mmol/kg)。

方法

磁共振多任务用于重建具有三个空间维度、一个用于动态 T 定量的 T 恢复维度和一个用于对比动力学的 DCE 维度的 5D 图像。使用双室交换模型从动态 T 图中估计动力学参数 、 、 、 。在 20 名健康受试者中评估了 LD-MT-DCE 的重复性和与标准剂量 MT-DCE 的一致性。在 7 例三阴性乳腺癌患者中,在同一成像会话中比较了 LD-MT-DCE 图像质量和诊断结果与标准剂量临床 DCE 的结果。采用单向不平衡方差分析和 Tukey 检验评估对照组和患者组之间动力学参数的统计学意义。

结果

LD-MT-DCE 技术重复性好,与标准剂量 MT-DCE 一致,图像质量优异。LD-MT-DCE 的诊断结果与临床结果吻合良好。 、 、 和 在恶性肿瘤与正常乳腺组织之间(P<0.001、<0.001 和<0.001),以及在恶性肿瘤与良性肿瘤之间(P=0.020、0.003 和<0.001)均有显著差异。

结论

LD-MT-DCE 技术重复性好,与标准剂量临床 DCE 相比,具有优异的图像质量和等效的诊断能力。估计的动力学参数能够区分正常乳腺组织与良性和恶性肿瘤。