Krischik Dirk, Tokgöz Selen Ergin, van Orten Andreas, Friedmann Anton, Bilhan Hakan
Private Dental Practice Zahnärzte Do24, Dortmunder Str. 24-28, 45731 Waltrop, Germany.
Private Dental Practice, Dentaglobal Oral Health Centre, Adalet Str 2131/10 No:3A, 35530 Bayraklı/İzmir, Turkey.
Dent J (Basel). 2021 Jan 8;9(1):5. doi: 10.3390/dj9010005.
The achievement of sufficient implant stability in poor quality bone seems to be a challenge. Most manufacturers develop special dental implants, which are claimed to show higher stability even in very soft bone. The aim of this experimental study was to compare two recently introduced dental implants with differing thread designs. A total of 11 implants of each group were inserted in the part of the fresh bovine ribs, corresponding to very soft bone. The primary stability was measured with resonance frequency analysis (RFA) and Periotest; the average of two measurements for each method and for each implant was taken and statistical analysis was applied. The highest stability values were obtained with the ICX Active Master implants, followed by the Conelog Progressive-Line implants placed with the very soft bone protocol. The primary stability values of the Conelog Progressive-Line implants inserted by the very soft bone protocol and the ICX Active Master implants placed with the standard protocol showed sufficient stability for immediate loading in low-density bone. Within the limitations of this study, the thread design of the implants and underdimensioned implant bed preparation seem to be effective for better primary stability in cancellous bone.
在质量较差的骨组织中实现足够的种植体稳定性似乎是一项挑战。大多数制造商开发了特殊的牙科种植体,据称这些种植体即使在非常松软的骨组织中也能表现出更高的稳定性。本实验研究的目的是比较两种最近推出的具有不同螺纹设计的牙科种植体。每组共11枚种植体被植入新鲜牛肋骨中相当于非常松软骨组织的部位。使用共振频率分析(RFA)和牙周探针测量初始稳定性;记录每种方法和每个种植体两次测量的平均值并进行统计分析。ICX Active Master种植体获得了最高的稳定性值,其次是按照非常松软骨组织植入方案植入的Conelog Progressive-Line种植体。按照非常松软骨组织植入方案植入的Conelog Progressive-Line种植体以及按照标准植入方案植入的ICX Active Master种植体的初始稳定性值表明,在低密度骨中即刻负重具有足够的稳定性。在本研究的局限性范围内,种植体的螺纹设计和种植窝预备不足似乎对提高松质骨中的初始稳定性有效。