Filters for radiation reduction: a comparison.

作者信息

Kohn M L, Gooch A W, Keller W S

机构信息

Department of Medical Physics, Mercy Hospital, Scranton, PA 18501.

出版信息

Radiology. 1988 Apr;167(1):255-7. doi: 10.1148/radiology.167.1.3347728.

Abstract

Aluminum, yttrium, copper, and R-filters were evaluated with respect to effective beam energy, entrance radiation exposure to the skin, x-ray tube loading, and the quality of resultant radiographs. Since all filters yielded acceptable radiographs, a figure of merit was developed to determine which filter delivered the lowest entrance radiation exposure for the lowest increase in x-ray tube loading. The figure of merit indicated that the best results were obtained with the 0.1-mm Cu filter with 2 mm Al at 60-105 kVp and with the 0.2-mm Cu filter with 2 mm Al at 105-120 kVp. Cu filters produced radiographs with quality comparable to that of R- or Y filters. Cu filters are inexpensive alternatives to rare-earth filters and yielded a reduction in radiation exposure with less x-ray tube loading.