Gjermeni Erind, Doering Michael, Richter Sergio, Hindricks Gerhard, Bode Kerstin
Department of Electrophysiology, Heart Center Leipzig at University of Leipzig, Leipzig, Germany.
Department of Electrophysiology, Heart Center Leipzig at University of Leipzig, Leipzig, Germany.
JACC Clin Electrophysiol. 2021 Jan;7(1):1-5. doi: 10.1016/j.jacep.2020.10.006.
Cardiovascular implantable electronic devices can initiate and sustain pacemaker-mediated arrhythmias. Endless loop tachycardia and repetitive non-re-entrant ventriculoatrial synchrony (RNRVAS) are well-described examples of pacemaker-mediated arrhythmias. However, such arrhythmias only occur in the presence of ventriculoatrial conduction. We identified a novel pacemaker-mediated arrhythmia that closely mimics RNRVAS but in the absence of ventriculoatrial conduction. We identified these arrhythmias in 9 patients, all with a St. Jude Medical/Abbott device, recorded as inappropriate mode-switch episodes by a device algorithm that includes paced events in the atrial counter. This report describes the mechanism, discusses clinical implications, and outlines programming options to eliminate this pseudo-RNRVAS.