EuroIntervention. 2021 Nov 19;17(10):856-864. doi: 10.4244/EIJ-D-20-00697.
THV implantation within failed surgical valves is well established. However, the implications of THV implantation within failed THVs are poorly understood.
This study aimed to assess the impact of different transcatheter heart valve (THV) designs and implant positioning strategies on hydrodynamic performance after redo transcatheter aortic valve implantation (TAVI).
THVs of varying design (SAPIEN 3, Evolut PRO, ACURATE neo, ALLEGRA, and Portico) and size were implanted inside SAPIEN XT and Evolut R THVs. Hydrodynamic function as per International Organization for Standardization (ISO) specifications was evaluated using a pulse duplicator, and multi-modality imaging was performed.
The majority of tested THV-in-THV combinations resulted in stable anchoring of the new implant. However, some combinations resulted in unstable anchoring with the potential for dislodgement or embolisation. Hydrodynamic function was favourable for all tested THV designs implanted in the intra-annular SAPIEN XT THV. SAPIEN 3 implantation within an Evolut THV with supra-annular leaflets must be appropriately sized to ensure adequate anchoring. Avoidance of an intra-annular deployment mitigated leaflet overhang of the Evolut leaflets and optimised hydrodynamic performance.
This study demonstrates that most THV designs and implantation strategies can result in favourable hydrodynamic performance following redo TAVI. Whether the leaflets of a failed THV are intra- or supra-annular may have important implications for the positioning of a redo-THV implant. Certain THV designs or implantation positions may be more desirable when performing redo TAVI.
经导管心脏瓣膜(THV)植入术治疗失败后的再次植入已得到广泛认可。然而,THV 植入术治疗失败后的相关影响尚未被充分认识。
本研究旨在评估不同经导管心脏瓣膜(THV)设计和植入定位策略对再次行经导管主动脉瓣置换术(TAVI)后血流动力学性能的影响。
将不同设计(SAPIEN 3、Evolut PRO、ACURATE neo、ALLEGRA 和 Portico)和大小的 THV 植入 SAPIEN XT 和 Evolut R THV 内。使用脉动复制仪按照国际标准化组织(ISO)规范评估血流动力学功能,并进行多模态成像。
大多数测试的 THV-in-THV 组合均导致新植入物稳定固定。然而,某些组合导致固定不稳定,存在移位或栓塞的潜在风险。在植入 SAPIEN XT THV 的环形内的所有测试的 THV 设计中,血流动力学功能均良好。在具有瓣上叶的 Evolut THV 中植入 SAPIEN 3 时,必须适当选择植入物尺寸以确保充分固定。避免环形内植入可减轻 Evolut 瓣叶的瓣叶悬垂,并优化血流动力学性能。
本研究表明,大多数 THV 设计和植入策略在再次行经 TAVI 后可获得良好的血流动力学性能。失败的 THV 的瓣叶是环形内还是瓣上可能对再次植入 THV 植入物的定位具有重要影响。在进行再次行经 TAVI 时,某些 THV 设计或植入位置可能更可取。