Abstract

PURPOSE

To evaluate the characteristics of needle artifacts and the accuracy of needle insertion using a 1.2 Tesla open magnetic resonance imaging (MRI) system in a phantom.

MATERIALS AND METHODS

First, the apparent width of the needle on the MRI and the needle tip position error of 16- and 18-gauge MRI-compatible introducer needles and a 17-gauge cryoneedle were examined with different needle angles (0°, 30°, 45°, 60°, and 90°) to the main magnetic field (B0), sequence types (balanced steady-state acquisition with rewound gradient echo [BASG] and T2-weighted fast spin echo [FSE] sequence), and frequency encoding directions. Second, the accuracy of needle insertion was evaluated after 10 MRI fluoroscopy-guided insertions in a phantom.

RESULTS

The apparent needle widths was larger when the angle of the needle axis relative to B0 was larger. The needles appeared larger on BASG than on T2-weighted FSE images, with the largest apparent widths of 16-, 17-, and 18-gauge needles of 14.3, 11.6, and 11.0mm, respectively. The apparent needle tip position was always more distal than the actual position on BASG images, with the largest longitudinal error of 4.0mm. Meanwhile, the 16- and 18-gauge needle tips appeared more proximal on T2-weighted FSE images with right-to-left frequency encoding direction. The mean accuracy of MRI fluoroscopy-guided needle insertion was 3.1mm.

CONCLUSION

These experiments clarify the characteristics of needle artifacts in a 1.2 Tesla open MRI. With this system, the MRI fluoroscopy-guided needle insertion demonstrated an acceptable accuracy for clinical use.

摘要

目的

在体模中评估使用 1.2T 开放式磁共振成像(MRI)系统的针状物伪影特征和针插入准确性。

材料和方法

首先,以不同的针轴相对于主磁场(B0)的角度(0°、30°、45°、60°和 90°)、序列类型(带重绕梯度回波的平衡稳态采集[BASG]和 T2 加权快速自旋回波[T2W-FSE]序列)和频率编码方向,检查了 16 号、18 号和 17 号 MRI 兼容导入器针以及 17 号冷冻针的 MRI 上的针显宽和针尖位置误差。其次,在体模中进行了 10 次 MRI 透视引导插入后,评估了针插入的准确性。

结果

针轴相对于 B0 的角度越大,针的显宽越大。BASG 图像上的针比 T2W-FSE 图像上的针显宽更大,16 号、17 号和 18 号针的显宽分别为 14.3、11.6 和 11.0mm。BASG 图像上的针尖位置总是比实际位置更靠后,最大纵向误差为 4.0mm。同时,T2W-FSE 图像上的右侧到左侧频率编码方向上,16 号和 18 号针的针尖看起来更靠前。MRI 透视引导针插入的平均准确性为 3.1mm。

结论

这些实验阐明了 1.2T 开放式 MRI 中的针状物伪影特征。使用该系统,MRI 透视引导的针插入具有临床应用的可接受准确性。