Suppr超能文献

无金属套筒与含金属套筒的3D打印计算机辅助种植手术导板的体外比较

In Vitro Comparison between Metal Sleeve-Free and Metal Sleeve-Incorporated 3D-Printed Computer-Assisted Implant Surgical Guides.

作者信息

Oh Kyung Chul, Shim June-Sung, Park Ji-Man

机构信息

Department of Prosthodontics, Yonsei University College of Dentistry, Seoul 03722, Korea.

出版信息

Materials (Basel). 2021 Jan 29;14(3):615. doi: 10.3390/ma14030615.

Abstract

The present study aims to compare the accuracy of metal sleeve-free 3D-printed computer-assisted implant surgical guides (MSF group) ( = 10) with metal sleeve-incorporated 3D-printed computer-assisted implant surgical guides (MSI group) ( = 10). Implants of diameter 4.0 mm and 5.0 mm were placed in the left second premolars and bilateral first molars, respectively, using a fully guided system. Closed-form sleeves were used in teeth on the left and open-form sleeves on the right. The weight differences of the surgical guides before and after implant placement, and angular deviations before and after implant placement were measured. Weight differences were compared with Student's -tests and angular deviations with Mann-Whitney tests. Cross-sectional views of the insert parts were observed with a scanning electron microscope. Preoperative and postoperative weight differences between the two groups were not statistically significant ( = 0.821). In terms of angular deviations, those along the mesiodistal direction for the left second premolars were significantly lower in the MSF group ( = 0.006). However, those along the mesiodistal direction for the bilateral molars and those along the buccolingual direction for all teeth were not significantly different ( > 0.05). 3D-printed implant surgical guides without metal sleeve inserts enable accurate implant placement without exhausting the guide holes, rendering them feasible for fully guided implant placement.

摘要

本研究旨在比较无金属套筒的3D打印计算机辅助种植手术导板(MSF组)(n = 10)与带有金属套筒的3D打印计算机辅助种植手术导板(MSI组)(n = 10)的准确性。使用全引导系统,分别在左侧第二前磨牙和双侧第一磨牙中植入直径4.0 mm和5.0 mm的种植体。左侧牙齿使用封闭式套筒,右侧牙齿使用开放式套筒。测量种植体植入前后手术导板的重量差异以及种植体植入前后的角度偏差。重量差异采用Student's t检验进行比较,角度偏差采用Mann-Whitney检验进行比较。用扫描电子显微镜观察插入部分的横截面视图。两组术前和术后的重量差异无统计学意义(P = 0.821)。在角度偏差方面,MSF组左侧第二前磨牙沿近远中方向的角度偏差显著更低(P = 0.006)。然而,双侧磨牙沿近远中方向的角度偏差以及所有牙齿沿颊舌方向的角度偏差无显著差异(P > 0.05)。无金属套筒插入物的3D打印种植手术导板能够在不耗尽导孔的情况下实现准确的种植体植入,使其适用于全引导种植体植入。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/c65c/7866124/7de51097d2a4/materials-14-00615-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验