胆红素构象光异构体对钒酸盐氧化法测定直接胆红素的影响。

Effects of bilirubin configurational photoisomers on the measurement of direct bilirubin by the vanadate oxidation method.

机构信息

Department of Pediatrics, Faculty of Medicine, Kagawa University, Kagawa, Japan.

Department of Clinical Laboratory, Faculty of Medicine, Kagawa University, Kagawa, Japan.

出版信息

Ann Clin Biochem. 2021 Jul;58(4):311-317. doi: 10.1177/0004563221999068. Epub 2021 Mar 9.

Abstract

BACKGROUND

Direct-reacting bilirubin concentrations measured using vanadate chemical oxidation method do not exactly match the conjugated bilirubin concentration. One of the causes is the effect of bilirubin photoisomers. However, the quantitative evaluation of the effects of these photoisomers has not been sufficiently conducted. In particular, the influence of bilirubin configurational isomers on direct bilirubin is the most critical factor.

METHODS

Sixteen residual serum samples were used. For quantitative analysis based on the change in direct bilirubin and bilirubin configurational isomer, samples were irradiated via blue light-emitting diodes to suppress the production of bilirubin structural isomers. Total bilirubin and direct bilirubin concentrations were measured using the vanadate chemical oxidation method. Concentrations of 4Z,15Z-bilirubin IXα and its photoisomers were measured using high-performance liquid chromatography. The sum of 4Z,15E-bilirubin IXα and 4E,15Z-bilirubin IXα was notated as bilirubin configurational isomer, and the differences between the measured values of the irradiated and non-irradiated samples were calculated and notated as ΔDB and ΔBCI.

RESULTS

In non-irradiated and irradiated samples, total bilirubin and direct bilirubin concentrations were 10.73 mg/dL with significant a decrease to 10.60 mg/dL and 0.69 mg/dL with a significant increase to 0.78 mg/dL, while bilirubin configurational isomer values were 1.00 mg/dL and 1.52 mg/dL, respectively. The linear regression equation revealed a significant positive correlation of Y = 0.187X-0.006 between ΔDB (Y) and ΔBCI (X).

CONCLUSION

Applying the vanadate chemical oxidation method affected approximately 19% of the bilirubin configurational isomer concentration for direct bilirubin. Extreme caution is necessary when interpreting the measured values of samples indicative of unconjugated hyperbilirubinaemia.

摘要

背景

采用钒酸盐化学氧化法测定的直接反应胆红素浓度与结合胆红素浓度并不完全匹配。原因之一是胆红素光异构体的影响。然而,这些光异构体的定量评估尚未充分进行。特别是,胆红素构型异构体对直接胆红素的影响是最关键的因素。

方法

使用 16 份剩余血清样本。为了基于直接胆红素和胆红素构型异构体的变化进行定量分析,通过蓝光发光二极管照射来抑制胆红素结构异构体的产生。使用钒酸盐化学氧化法测定总胆红素和直接胆红素浓度。使用高效液相色谱法测定 4Z,15Z-胆红素 IXα及其光异构体的浓度。4Z,15E-胆红素 IXα和 4E,15Z-胆红素 IXα的总和被标记为胆红素构型异构体,计算并标记照射和未照射样品的测量值之间的差异为 ΔDB 和 ΔBCI。

结果

在未照射和照射的样本中,总胆红素和直接胆红素浓度分别从 10.73mg/dL 显著下降至 10.60mg/dL 和 0.69mg/dL,显著增加至 0.78mg/dL,而胆红素构型异构体浓度分别为 1.00mg/dL 和 1.52mg/dL。线性回归方程显示 ΔDB(Y)和 ΔBCI(X)之间存在显著正相关,Y=0.187X-0.006。

结论

应用钒酸盐化学氧化法会影响直接胆红素构型异构体浓度的约 19%。在解释未结合高胆红素血症样本的测量值时需要格外小心。