McCullagh P J, McAllister H G, McCaffrey P M, Fox J S, Taggart A J, Riddell J G
Department of Mental Health, Queen's University of Belfast, UK.
J Med Eng Technol. 1988 Jan-Feb;12(1):7-14. doi: 10.3109/03091908809030150.
An automated system for the measurement of peripheral blood flow using venous occlusion plethysmography based on the low-cost Apple II microcomputer, together with purpose-built compressor unit and data acquisition interface, has been developed. The computer performs the dual role of controlling the timing of inflation and deflation of the occluding cuffs and recording the resulting increase in limb circumference. Ten 8 s epochs of data are acquired and analysed per session. Flow rate is computed using a least squares fit between 0.5 and 4.0 s after cuff inflation, giving on-line indication of blood flow. Venous capacitance and digital systolic pressure may be measured using additional algorithms. The system has been used for the investigation of circulatory disorders and in the assessment of drugs acting on the peripheral circulation.