Pinto Jader Camilo, Torres Fernanda Ferrari Esteves, Santos-Junior Airton Oliveira, Duarte Marco Antonio Hungaro, Guerreiro-Tanomaru Juliane Maria, Tanomaru-Filho Mario
Department of Restorative Dentistry, School of Dentistry, São Paulo State University, São Paulo, Brazil.
Department of Dentistry, Endodontics and Dental Materials, School of Dentistry, University of São Paulo São Paulo, Brazil.
Eur J Dent. 2021 May;15(2):247-252. doi: 10.1055/s-0041-1723065. Epub 2021 Feb 23.
The aim of this study was to investigate the effect of additional apical preparation using the ProDesign Logic (PDL) 50/.01 rotary heat-treated nickel-titanium (NiTi) file with a larger diameter and minimal taper for retreatment of curved root canals.
Mesial curved root canals of 12 mandibular molars were prepared using PDL 25/.06 and filled using the continuous wave of condensation technique and AH Plus sealer. After retreatment using ProDesign S (PDS) 25/.08, PDL 25/.06 and PDL 35/05, a complementary procedure was performed with PDL 50/.01. Microcomputed tomography (micro-CT) scanning was performed before and after retreatment procedures. The cyclic fatigue resistance of unused PDS 25/.08, PDL 25/.06, PDL 35/.05 and PDL 50/.01 instruments ( = 12) was evaluated in a stainless-steel device.
Data on the volumes of the root canals and the remaining filling materials were submitted to the paired t-test. Cyclic fatigue resistance data was submitted to one-way ANOVA and Tukey's tests (α = 0.05).
Use of PDL 50/.01 decreased the remaining filling materials in the apical third ( 0.05). The root canal volume was similar in the cervical and middle thirds before and after preparation using PDL 50/.01 ( 0.05). PDL 50/.01 presented the highest cyclic fatigue resistance ( 0.05).
Use of the PDL 50/.01 instrument as an additional apical preparation for retreatment of curved root canals improved filling material removal in the apical third, while maintaining the dentin in the cervical and middle thirds. In addition, PDL 50/.01 presented high-flexural resistance.
本研究旨在探讨使用ProDesign Logic(PDL)50/.01旋转热处理镍钛(NiTi)锉进行额外的根尖预备对弯曲根管再治疗的效果,该锉具有更大的直径和最小的锥度。
使用PDL 25/.06对12颗下颌磨牙的近中弯曲根管进行预备,采用连续波热凝技术和AH Plus封闭剂进行充填。在使用ProDesign S(PDS)25/.08、PDL 25/.06和PDL 35/05进行再治疗后,使用PDL 50/.01进行补充操作。在再治疗前后进行微计算机断层扫描(micro-CT)。在不锈钢装置中评估未使用的PDS 25/.08、PDL 25/.06、PDL 35/.05和PDL 50/.01器械(n = 12)的抗循环疲劳性能。
将根管和剩余充填材料的体积数据进行配对t检验。抗循环疲劳性能数据进行单因素方差分析和Tukey检验(α = 0.05)。
使用PDL 50/.01减少了根尖三分之一处的剩余充填材料(P < 0.05)。在使用PDL 50/.01进行预备前后,根管在颈段和中段的体积相似(P > 0.05)。PDL 50/.01表现出最高的抗循环疲劳性能(P < 0.05)。
使用PDL 50/.01器械作为弯曲根管再治疗的额外根尖预备,可改善根尖三分之一处充填材料的去除,同时保留颈段和中段的牙本质。此外,PDL 50/.01具有较高的抗弯曲性能。