Lee C R, Esnaud H
Laboratoires d'Etudes et de Recherches Synthélabo (LERS), Bagneux, France.
Biomed Environ Mass Spectrom. 1988 Mar 1;15(5):249-52. doi: 10.1002/bms.1200150503.
Determination of the low concentrations (less than 10-100 pg ml-1) of melatonin in blood plasma by gas chromatography/mass spectrometry has previously required the use of negative ionization (electron capture) mass spectrometry, a technique particularly delicate to set up. We show that a general-purpose medium-resolution instrument gives adequate sensitivity in electron impact mode, when used with a capillary gas chromatographic column-switching device. A straightforward extraction procedure gives good specificity, and the limit of useful measurement is better than 10 pg per 1 ml sample.
通过气相色谱/质谱法测定血浆中低浓度(低于10 - 100 pg/ml)的褪黑素,以前需要使用负离子(电子捕获)质谱法,这是一种设置起来特别精细的技术。我们表明,当与毛细管气相色谱柱切换装置一起使用时,通用的中分辨率仪器在电子轰击模式下具有足够的灵敏度。一种简单的提取程序具有良好的特异性,有用测量限优于每1 ml样品10 pg。