Aquili Alberto, Canè Pier Paolo, Fravisini Marco, Farinelli Luca, Procaccini Roberto, Gigante Antonio Pompilio
Clinical Orthopaedics, Università Politecnica delle Marche, Ancona, Italy.
Centro di Artroscopia e Chirurgia del Ginocchio, Clinica "Sol et Salus", Rimini, Italy.
J Orthop. 2021 Feb 12;24:86-90. doi: 10.1016/j.jor.2021.02.007. eCollection 2021 Mar-Apr.
We assessed the accuracy of an extramedullary guide system for femoral component alignment in TKA.
We retrospectively analysed 87 total knee arthroplasties using a newly developed extramedullary guide system.
Correct postoperative coronal plane femoral component, with deviation from neutral alignment by 3° or less, was found in 87% of the study population. The percentages were 100%, 91% and 79% for HKA of 0-3°, 3-10° and >10°respectively.
The correct use of the extramedullary guide system allows the restoration of the neutral mechanical axes of the lower limb, especially in cases of limited varus deformity.
我们评估了一种用于全膝关节置换术(TKA)中股骨组件对线的髓外导向系统的准确性。
我们回顾性分析了87例使用新开发的髓外导向系统进行的全膝关节置换术。
在87%的研究人群中发现术后股骨组件在冠状面正确,与中立对线的偏差在3°或更小。对于髋膝角(HKA)为0 - 3°、3 - 10°和>10°的情况,百分比分别为100%、91%和79%。
正确使用髓外导向系统可恢复下肢的中立机械轴,尤其是在轻度内翻畸形的情况下。