Department of Orthopedic Surgery, Korea University College of Medicine, Guro Hospital, Seoul, South Korea.
Department of Orthopedic Surgery, Kyung Hee University Hospital, College of Medicine, Kyung-Hee University, Seoul, South Korea.
Sci Rep. 2021 Mar 10;11(1):5539. doi: 10.1038/s41598-021-85193-7.
This study aimed to evaluate (1) the overall reasons for first revision in CoC THAs; (2) whether the reasons for revision differ between third-generation and fourth-generation CoC THAs; and (3) the specific factors associated with bearing-related problems as the reason for revision. We retrospectively reviewed 2045 patients (2194 hips) who underwent first revision THA between 2004 and 2013, among which 146 hips with CoC bearings underwent revision. There were 92 hips with third-generation ceramic bearings and 54 hips with fourth-generation ceramic bearings. The major reasons for CoC THA revisions were ceramic fracture and loosening of the cup or stem. When ceramic fracture, squeaking, incorrect ceramic insertion, and unexplained pain were defined as directly related or potentially related to ceramic use, 28.8% (42/146) of CoC revisions were associated with bearing-related problems. Among the third-generation ceramic bearings, revision was performed in 41.3% (38/92) of cases owing to bearing-related problems whereas revisions were performed for only 7.4% (4/54) of cases with fourth-generation ceramic bearings owing to bearing-related problems (p < 0.001). Younger age, lower American Society of Anesthesiologists (ASA) grade, and preoperative diagnosis of osteonecrosis were factors related to CoC THA revisions due to bearing-related problems.
(1) CoC THR 初次翻修的总体原因;(2) 第三代和第四代 CoC THR 之间的翻修原因是否存在差异;(3) 与轴承相关问题作为翻修原因的具体因素。我们回顾性分析了 2004 年至 2013 年间接受初次翻修 THR 的 2045 例患者(2194 髋),其中 CoC 轴承翻修 146 髋。有 92 髋为第三代陶瓷轴承,54 髋为第四代陶瓷轴承。CoC THR 翻修的主要原因是陶瓷破裂和杯或柄松动。当陶瓷破裂、嘎吱作响、陶瓷插入不正确和不明原因的疼痛被定义为与陶瓷使用直接相关或潜在相关时,28.8%(42/146)的 CoC 翻修与轴承相关问题有关。在第三代陶瓷轴承中,由于轴承相关问题,有 41.3%(38/92)的病例需要进行翻修,而在第四代陶瓷轴承中,仅有 7.4%(4/54)的病例需要进行翻修,原因是轴承相关问题(p<0.001)。年龄较小、较低的美国麻醉医师协会(ASA)分级和术前诊断为骨坏死是与 CoC THR 因轴承相关问题翻修相关的因素。