Papaioannou Ioannis, Repantis Thomas, Korovessis Panagiotis
Orthopedic Department, General Hospital of Patras, Greece.
J Clin Orthop Trauma. 2020 Aug 13;13:92-94. doi: 10.1016/j.jcot.2020.08.006. eCollection 2021 Feb.
Gamma 3 nail is a wide spread intramedullary device for fixation of per trochanteric fractures. Cut out of the lag screw is the most common complication of this implant. We present a 62-year-old female patient, who underwent a total hip arthroplasty following cut out of a Gamma 3 nail in the femoral neck. The cause of the cut out in our case is actually unique. Our intraoperative findings accompanied with the radiographic evaluation argue that the malposition of the set screw was the cause of failure, due to the rotational instability of femoral head-lag screw unit. We present this case with detailed description, highlighting the proper use of this specific nail and appose a brief literature review.