Calibration for quantitative Fc-glycosylation analysis of therapeutic IgG1-type monoclonal antibodies by using glycopeptide standards.

作者信息

Cao Cuiyan, Yu Long, Zhang Xiuli, Dong Xuefang, Yuan Jingli, Liang Xinmiao

机构信息

Key Laboratory of Separation Science for Analytical Chemistry, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian, 116023, China; University of Chinese Academy of Sciences, Beijing, 100049, China; State Key Laboratory of Fine Chemicals, School of Chemistry, Dalian University of Technology, Dalian, 116024, China.

Key Laboratory of Separation Science for Analytical Chemistry, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian, 116023, China.

出版信息

Anal Chim Acta. 2021 Apr 15;1154:338306. doi: 10.1016/j.aca.2021.338306. Epub 2021 Feb 12.

Abstract

Fc-glycosylation has crucial impact on the efficacy and safety of IgG-type therapeutic monoclonal antibodies (mAbs). In order to enhance the performance of MS-based bottom-up quantitation strategy, a library of glycopeptide standards containing 26 common IgG1-type Fc-glycoforms has been constructed via modified two-dimensional hydrophilic interaction liquid chromatography (HILIC) purification. Taking advantage of the acquired glycopeptide standards, calibrated quantitation strategy for Fc-glycosylation analysis of mAbs was established and evaluated on the basis of three LC-MS-based methods, including HILIC-MRM (multiple reaction monitoring), HILIC-SIM (selected ion monitor) and RPLC-SIM. Molar concentrations of eleven individual Fc-glycoforms (0.03 ± 0.001-13.77 ± 0.64 nmol mg) as well as degree of fucosylation (75.44-97.04%), galactosylation (3.39-49.47%) and mannosylation (1.12-21.22%) in six IgG1-type mAbs were achieved. In addition, Fc-glycosylation site occupancy was also determined from 98.05% to 99.83%. Compared with traditional MS-based quantitation via peak area normalization, the quantitation accuracy and precision of the calibrated strategy had been remarkably improved, especially when combining with HILIC separation. In addition, the transferability of calibrated quantitation as assessed by using MRM-based method had also been significantly enhanced on different instruments from different laboratories. This calibrated quantitation strategy using glycopeptide standards as calibrators will be useful for Fc-glycosylation analysis of IgG1-type mAbs with multiple glycosylation sites.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验