Theivacumar Nadarajah Selva, Qureshi Mahim I, Glasgow Simon, Najem Mojahid
West London Vascular and Interventional Centre, Northwick Park Hospital, Harrow, United Kingdom.
J Vasc Surg Cases Innov Tech. 2021 Jan 28;7(1):183-187. doi: 10.1016/j.jvscit.2020.12.014. eCollection 2021 Mar.
Suture-based vascular closure devices are used in percutaneous endovascular procedures. However, failures are not uncommon. We have described our initial experience with two adjunct techniques to reinforce the suture-based vascular closure device (ProGlide; Abbot Vascular, Santa Clara, Calif) after percutaneous endovascular aneurysm repair. The threads of the ProGlide device (Abbot Vascular) were passed through a pledget with the help of a needle, which was secured to the puncture site to allow for traction compression. The use of the techniques can be helpful if the suture-based vascular closure devices fail to achieve immediate and complete hemostasis. The use of these adjuncts could reduce the incidence of closure-related complications after percutaneous endovascular procedures.
基于缝线的血管闭合装置用于经皮血管腔内手术。然而,失败情况并不罕见。我们描述了在经皮血管腔内动脉瘤修复术后,使用两种辅助技术来加强基于缝线的血管闭合装置(ProGlide;雅培血管,加利福尼亚州圣克拉拉)的初步经验。在针的帮助下,将ProGlide装置(雅培血管)的缝线穿过一个小垫片,该小垫片固定在穿刺部位以进行牵引压迫。如果基于缝线的血管闭合装置未能实现即时和完全止血,使用这些技术可能会有所帮助。使用这些辅助手段可以降低经皮血管腔内手术后与闭合相关并发症的发生率。