BMC Med Imaging. 2021 Mar 23;21(1):59. doi: 10.1186/s12880-021-00589-1.
This study was performed to determine whether in-laboratory specimen radiography reduces turnaround time or block utilization in surgical pathology.
Specimens processed during a 48-day trial of an in-lab cabinet radiography device (Faxitron) were compared to a control group of specimens imaged in the mammography suite during a prior 1-year period, and to a second group of specimens not undergoing imaging of any type.
Cases imaged in the mammography suite had longer turnaround time than cases not requiring imaging (by 1.15 days for core biopsies, and 1.73 days for mastectomies; p < 0.0001). In contrast, cases imaged in-lab had turnaround time that was no longer than unimaged cases (p > 0.05 for core biopsies, lumpectomies and mastectomies). Mastectomies imaged in-lab required submission of fewer blocks than controls not undergoing any imaging (mean reduction of 10.6 blocks).
Availability of in-lab radiography resulted in clinically meaningful improvements in turnaround time and economically meaningful reductions in block utilization.