Kalender W A, Brestowsky H, Felsenberg D
Siemens Medical Systems, Erlangen, Federal Republic of Germany.
Radiology. 1988 Jul;168(1):219-21. doi: 10.1148/radiology.168.1.3380963.
The selection of the computed tomographic (CT) section can introduce errors in reproducibility of several percent for bone mineral measurements. An algorithm for automated determination of the midvertebral CT section from survey radiographs was developed; it is based on ridge-line tracking of the vertebral end plates. The midvertebral section is defined by all points with equal perpendicular distance to the regression lines fitted through these end-plate contours. The algorithm can be corrected by manual interaction in case of failure. It was tested on T-12 to L-4 vertebrae on 25 randomly selected localizer radiographs. Of the 250 endplate contours, 223 (89.2%) were tracked automatically; the remaining 27 (10.8%) required user interaction for successful completion.
计算机断层扫描(CT)切片的选择可能会在骨矿物质测量的可重复性方面引入百分之几的误差。开发了一种从定位X光片中自动确定椎体中部CT切片的算法;该算法基于椎体终板的脊线跟踪。椎体中部切片由与拟合这些终板轮廓的回归线具有相等垂直距离的所有点定义。在出现故障的情况下,该算法可通过人工交互进行校正。在25张随机选择的定位X光片上对T-12至L-4椎体进行了测试。在250个终板轮廓中,223个(89.2%)被自动跟踪;其余27个(10.8%)需要用户交互才能成功完成。