Lead perforation is one of the major complications of pacemaker implantation, but cases of right ventricular (RV) lead perforation through the septum and left ventricle are rarely reported. We described a rare case of left ventricular (LV) free wall perforation by an RV lead and the management of this complication.
An 84-year-old man was admitted with a dual-chamber pacemaker due to pacing failure caused by an RV lead fracture. New lead implantation was performed on the next day, but pacing failure occurred again on the second post-operative day (POD). We found the lead perforation on the fluoroscopy during temporary pacemaker insertion. Computed tomography scan and transthoracic echocardiogram showed that the added lead perforated through both the septum and LV free wall. A new lead was inserted on the fourth POD, and an off-pump open chest surgery for extraction of the penetrating lead was performed uneventfully on the 20th POD.
We considered that some features of the lead (SelectSecure 3830-69, Medtronic) may be related to this complication, as the lead was very thin, had a non-retractable bare screw and was inserted with a dedicated delivery catheter. We have to be careful when performing implantation of this kind of lead to avoid such a rare complication.
导线穿孔是起搏器植入的主要并发症之一,但右心室(RV)导线经间隔和左心室穿孔的病例鲜有报道。我们描述了一例罕见的右心室导线致左心室游离壁穿孔病例及该并发症的处理。
一名84岁男性因右心室导线断裂导致起搏失败而植入双腔起搏器入院。次日进行了新导线植入,但术后第二天(POD)再次出现起搏失败。在临时起搏器植入期间的荧光透视检查中发现导线穿孔。计算机断层扫描和经胸超声心动图显示,新增导线穿透了间隔和左心室游离壁。在术后第四天植入了一根新导线,并于术后第20天顺利进行了非体外循环开胸手术取出穿透的导线。
我们认为导线(美敦力SelectSecure 3830 - 69)的一些特性可能与该并发症有关,因为该导线非常细,有不可回缩的裸螺旋且通过专用输送导管插入。植入此类导线时我们必须小心,以避免这种罕见并发症。