Department of Artificial Organs, National Cerebral and Cardiovascular Center Research Institute, 6-1 Kishibe-Shimmachi, Suita, Osaka, 564-8565, Japan.
Department of Cardiac Surgery, University of Tokyo, Tokyo, Japan.
J Artif Organs. 2021 Dec;24(4):419-424. doi: 10.1007/s10047-021-01265-5. Epub 2021 Apr 6.
Although the influence of continuous-flow left ventricular assist device (CF-LVAD) support on peripheral circulation has been widely discussed, its monitoring modalities are limited. The aim of this study was to assess the peripheral circulation using the laser speckle flowgraph (LSFG) which can quantitatively measure the ocular blood flow. We implanted a centrifugal CF-LVAD (EVAHEART; Sun Medical Technology Research Corporation, Nagano, Japan) in five adult goats (body weight 44.5 ± 2.9 kg) under general anesthesia. The waveform of the central retinal artery using the mean blur rate (MBR) for ocular blood velocity and fluctuations as a parameter of pulsatility were obtained before LVAD implantation and after LVAD full-bypass support. The MBR waveform and LSFG fluctuation data were compared with the waveform and pulsatility index of the external carotid artery using an ultrasonic flow meter to evaluate circulatory patterns at different levels. The MBR waveform pattern of the central retinal artery was pulsatile before LVAD implantation and less pulsatile under LVAD full bypass. The fluctuation was 14.7 ± 1.86 before LVAD implantation and 3.85 ± 0.61 under LVAD full bypass (p < 0.01), respectively. The fluctuations of LSFG showed a strong correlation with the pulsatility index of the external carotid artery meaning that similar changes in circulatory pattern were observed at two different levels. Measuring the ocular blood flow using LSFG has potential utility for the assessment of the status of the peripheral circulation and its pulsatility during CF-LVAD.
虽然连续血流左心室辅助装置(CF-LVAD)支持对周围循环的影响已被广泛讨论,但对其监测方式仍有限制。本研究旨在使用激光散斑血流图(LSFG)评估外周循环,该方法可定量测量眼部血流。我们在五只成年山羊(体重 44.5±2.9kg)全麻下植入离心式 CF-LVAD(EVAHEART;日本长野县 Sun Medical Technology Research Corporation)。在植入 LVAD 之前和完全旁路支持之后,使用中央视网膜动脉的平均模糊率(MBR)作为眼血流速度和波动的参数,获得眼部血流的波形。将 MBR 波形和 LSFG 波动数据与超声流量计评估不同水平的循环模式时的颈外动脉的波形和搏动指数进行比较。在植入 LVAD 之前,中央视网膜动脉的 MBR 波形呈搏动性,而在 LVAD 完全旁路时搏动性较弱。波动分别为植入 LVAD 之前的 14.7±1.86 和 LVAD 完全旁路时的 3.85±0.61(p<0.01)。LSFG 的波动与颈外动脉的搏动指数呈强相关,这意味着在两个不同水平上观察到相似的循环模式变化。使用 LSFG 测量眼部血流具有评估 CF-LVAD 期间外周循环及其搏动状态的潜在应用价值。