Influence of CBCT metal artifact reduction on vertical radicular fracture detection.

作者信息

Oliveira Mariana Rodrigues, Sousa Thiago Oliveira, Caetano Aline Ferreira, de Paiva Rogério Ribeiro, Valladares-Neto José, Yamamoto-Silva Fernanda Paula, Silva Maria Alves Garcia

机构信息

Department of Stomatological Sciences, School of Dentistry, Federal University of Goiás, Goiânia, Goiás, Brazil.

Oral Radiology Center - CIRO, Goiânia, Goiás, Brazil.

出版信息

Imaging Sci Dent. 2021 Mar;51(1):55-62. doi: 10.5624/isd.20200191. Epub 2021 Jan 29.

Abstract

PURPOSE

This study evaluated the influence of a metal artifact reduction (MAR) tool in a cone-beam computed tomography (CBCT) device on the diagnosis of vertical root fractures (VRFs) in teeth with different root filling materials.

MATERIALS AND METHODS

Forty-five extracted human premolars were classified into three subgroups; 1) no filling; 2) gutta-percha; and 3) metallic post. CBCT images were acquired using an Orthopantomograph 300 unit with and without a MAR tool. Subsequently, the same teeth were fractured, and new CBCT scans were obtained with and without MAR. Two oral radiologists evaluated the images regarding the presence or absence of VRF. Receiver operating characteristic (ROC) curves and diagnostic tests were performed.

RESULTS

The overall area under the curve values were 0.695 for CBCT with MAR and 0.789 for CBCT without MAR. The MAR tool negatively influenced the overall diagnosis of VRFs in all tested subgroups, with lower accuracy (0.45-0.72), sensitivity (0.6-0.67), and specificity (0.23-0.8) than were found for the images without MAR. In the latter group, the accuracy, sensitivity, and specificity values were 0.68-0.77, 0.67-083, and 0.53-087, respectively. However, no significant difference was found between images with and without MAR for the no filling and gutta-percha subgroups (>0.05). In the metallic post subgroup, CBCT showed a significant difference according to MAR use (<0.05).

CONCLUSION

The OP 300 MAR tool negatively influenced the detection of VRFs in teeth with no root canal filling, gutta-percha, or metallic posts. Teeth with metallic posts suffered the most from the negative impact of MAR.

摘要

目的

本研究评估了锥形束计算机断层扫描(CBCT)设备中的金属伪影减少(MAR)工具对不同根管充填材料牙齿垂直根折(VRF)诊断的影响。

材料与方法

45颗拔除的人类前磨牙分为三个亚组;1)未充填;2)牙胶;3)金属桩。使用带有和不带有MAR工具的Orthopantomograph 300设备获取CBCT图像。随后,将相同的牙齿折断,并在有和没有MAR的情况下获得新的CBCT扫描图像。两位口腔放射科医生评估图像中是否存在VRF。进行了受试者操作特征(ROC)曲线和诊断测试。

结果

带有MAR的CBCT曲线下总面积值为0.695,不带有MAR的CBCT为0.789。MAR工具对所有测试亚组中VRF的总体诊断产生负面影响,其准确性(0.45 - 0.72)、敏感性(0.6 - 0.67)和特异性(0.23 - 0.8)均低于不带有MAR的图像。在后者组中,准确性、敏感性和特异性值分别为0.68 - 0.77、0.67 - 0.83和0.53 - 0.87。然而,未充填和牙胶亚组中带有和不带有MAR的图像之间未发现显著差异(>0.05)。在金属桩亚组中,根据是否使用MAR,CBCT显示出显著差异(<0.05)。

结论

OP 300 MAR工具对无根管充填、牙胶或金属桩的牙齿中VRF的检测产生负面影响。带有金属桩的牙齿受MAR负面影响最大。