Burlina A, Bonvicini P, Plebani M, Zaninotto M
Department of Clinical Chemistry and Clinical Microscopy, University of Padua, Italy.
Clin Chim Acta. 1988 Mar 31;173(1):35-41. doi: 10.1016/0009-8981(88)90355-5.
The use of recommended methods has very much improved the precision and accuracy of enzymatic determinations. The use of 'enzyme reference materials' allows us to overcome the remaining difficulties. The possibility of extending standardization by using calibration materials to convert results obtained by different methods requires an essential property: 'commutability'. The relationship between patients' specimens and reference materials are conditioned by analytical causes (relative to the used methods) and clinical causes (relative to the condition of the human samples to be examined). Among the clinical causes, the different isoenzyme content of patients sera, induced by different pathologies, was considered. The effect of this influence has been investigated with an analytical approach of the isoenzymes in the individual clinical samples. The results obtained on determination of alkaline phosphatase (ALP), creatine kinase (CK), and lactate dehydrogenase (LD), are discussed according to a model of investigation which can be utilized for the experimental protocols in the assessment of commutability.
推荐方法的使用极大地提高了酶促测定的精密度和准确度。“酶参考物质”的使用使我们能够克服剩余的困难。通过使用校准物质来扩展标准化以转换不同方法获得的结果,这需要一个基本特性:“互换性”。患者样本与参考物质之间的关系受分析原因(相对于所使用的方法)和临床原因(相对于待检查人体样本的状况)的制约。在临床原因中,考虑了由不同病理引起的患者血清中不同同工酶含量的情况。已通过对个体临床样本中的同工酶进行分析的方法研究了这种影响的效果。根据一种可用于评估互换性实验方案的研究模型,讨论了碱性磷酸酶(ALP)、肌酸激酶(CK)和乳酸脱氢酶(LD)测定所获得的结果。