双功超声检查在检测使用血管封堵器(Angio-Seal)过程中胶原海绵误置入动脉腔的应用:一例病例报告

Usefulness of duplex ultrasonography to detect collagen sponge misplacement into the arterial lumen during the use of Angio-Seal: A case report.

作者信息

Sugawara Masayo, Hoshina Katsuyuki, Chigira Mayumi, Suzuki Hirohide, Daimon Masao, Takayama Toshio, Ono Yoshikazu, Yatomi Yutaka

机构信息

Department of Clinical Laboratory Medicine, The University of Tokyo, Tokyo, Japan.

Department of Vascular Surgery, The University of Tokyo, Tokyo, Japan.

出版信息

SAGE Open Med Case Rep. 2021 Mar 30;9:2050313X211004818. doi: 10.1177/2050313X211004818. eCollection 2021.

Abstract

A 76-year-old man who had undergone percutaneous coronary intervention suffered intermittent claudication. Duplex ultrasonography revealed superficial femoral artery stenosis with an intraluminal heterogeneous echogenic mass. We suspected that stenosis was caused by the puncture procedure of the Angio-Seal. Open surgery revealed that a collagen sponge that should have been outside the arterial wall was misplaced in the wall with massive granulation, and atherectomy with patchplasty was performed. Vessel deterioration was considered due to several factors, including inappropriate access site, arterial wall calcification, and comorbidities like Behçet's disease. Ultrasonography is a convenient and useful method to evaluate arterial lesions.

摘要

一名接受过经皮冠状动脉介入治疗的76岁男性出现间歇性跛行。双功超声检查显示股浅动脉狭窄,管腔内有不均匀回声团块。我们怀疑狭窄是由血管封堵器的穿刺操作引起的。开放手术发现,本应位于动脉壁外的胶原海绵错位至动脉壁内,并伴有大量肉芽组织形成,遂行斑块旋切术及补片成形术。血管病变被认为是由多种因素导致的,包括穿刺部位不当、动脉壁钙化以及白塞病等合并症。超声检查是评估动脉病变的一种便捷且有用的方法。