Suppr超能文献

使用双能量计算机断层扫描对碘进行定量的准确性,重点关注低浓度。

Accuracy of iodine quantification using dual-energy computed tomography with focus on low concentrations.

机构信息

Department of Radiology, Uppsala University Hospital, Sweden.

Pharmeteus AB, Uppsala, Sweden.

出版信息

Acta Radiol. 2022 May;63(5):623-631. doi: 10.1177/02841851211009462. Epub 2021 Apr 22.

Abstract

BACKGROUND

Iodine quantification using dual-energy computed tomography (DECT) is helpful in characterizing, and follow-up after treatment of tumors. Some malignant masses, for instance papillary renal cell carcinomas (p-RCC), are hard to differentiate from benign lesions because of very low contrast enhancement. In these cases, iodine concentrations might be very low, and it is therefore important that iodine quantification is reliable even at low concentrations if this technique is used.

PURPOSE

To examine the accuracy of iodine quantification and to determine whether it is also accurate for low iodine concentrations.

MATERIAL AND METHODS

Twenty-six syringes with different iodine concentrations (0-30 mg I/mL) were scanned in a phantom model using a DECT scanner with two different kilovoltage and image reconstruction settings. Iodine concentrations were measured and compared to known concentration. Absolute and relative errors were calculated.

RESULTS

For concentrations of 1 mg I/mL or higher, there was an excellent correlation between true and measured iodine concentrations for all settings (R = 0.999-1.000;  < 0.001). For concentrations <1.0 mg I/mL, the relative error was greater. Absolute and relative errors were smaller using tube voltages of 80/Sn140 kV than 100/Sn140 kV ( < 0.01). Reconstructions using a 3.0-mm slice thickness had less variance between repeated acquisitions versus 0.6 mm ( < 0.001).

CONCLUSION

Iodine quantification using DECT was in general very accurate, but for concentrations < 1.0 mg I/mL the technique was less reliable. Using a tube voltage with larger spectral separation was more accurate and the result was more reproducible using thicker image reconstructions.

摘要

背景

使用双能计算机断层扫描(DECT)进行碘定量有助于对肿瘤进行特征描述和治疗后随访。某些恶性肿块,例如乳头状肾细胞癌(p-RCC),由于对比度增强非常低,很难与良性病变区分。在这些情况下,碘浓度可能非常低,如果使用该技术,则即使在低浓度下,碘定量也必须可靠。

目的

检查碘定量的准确性,并确定在低碘浓度下是否也准确。

材料和方法

使用具有两种不同千伏和图像重建设置的 DECT 扫描仪,在体模模型中对 26 个具有不同碘浓度(0-30mgI/mL)的注射器进行扫描。测量碘浓度并与已知浓度进行比较。计算绝对误差和相对误差。

结果

对于 1mgI/mL 或更高的浓度,对于所有设置,真实碘浓度与测量碘浓度之间均存在极好的相关性(R=0.999-1.000; <0.001)。对于浓度<1.0mgI/mL,相对误差更大。与 100/Sn140kV 相比,使用 80/Sn140kV 管电压的绝对误差和相对误差更小( <0.01)。与 0.6mm 相比,使用 3.0mm 切片厚度进行重建时,重复采集之间的差异更小( <0.001)。

结论

使用 DECT 进行碘定量通常非常准确,但对于浓度<1.0mgI/mL 的技术不太可靠。使用具有更大光谱分离的管电压更准确,并且使用更厚的图像重建结果更具可重复性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验