Wyantuti Santhy, Pratomo Uji, Manullang Lastri A, Hendrati Diana, Hartati Yeni Wahyuni, Bahti Husein H
Department of Chemistry, Faculty of Mathematics and Natural Sciences, Universitas Padjadjaran, Jl. Raya Bandung-Sumedang Km. 21, Jatinangor, Sumedang, 45363 Indonesia.
Heliyon. 2021 Apr 5;7(4):e06602. doi: 10.1016/j.heliyon.2021.e06602. eCollection 2021 Apr.
The development of methods for the efficient and reliable separation and routine analysis of rare-earth elements (REEs), including samarium (Sm), proceeds to draw in the interest of the many researchers, attributable to the similar physical and chemical properties of these elements. Note that although the voltammetric determination of Sm has been described in the literature, thus far, no chemometric and voltammetric methods for the quantification of the element in its mixtures with other lanthanides in an acetonitrile solution have been reported. This work was aimed toward the advancement of a method for the detection of Sm in acetonitrile, the intended function of which was to obtain a selective current response of Sm by Differential Pulse Voltammetry, utilizing the Box-Behnken experimental design, to identify the best conditions for the determination. In particular, the three selected factors for the experiment, namely the potential range, amplitude modulation, and the deposition time, were found to have optimal conditions of -1.5 to +1.0 V, 0.075 V, and 60 s, consecutively. The optimal conditions were observed to result in a selective current response for samarium with a detection limit of 2.25 mg/L and a limit of quantitation of 7.50 mg/L. Furthermore, the accuracy was 98.70% and the precision was found to be 1.91% in relative standard deviation (RSD), while the recovery was found to be 98.70%.
稀土元素(REEs),包括钐(Sm),高效可靠的分离方法及常规分析方法的开发一直吸引着众多研究人员的关注,这是由于这些元素具有相似的物理和化学性质。需要注意的是,尽管文献中已描述了钐的伏安测定法,但迄今为止,尚未有关于在乙腈溶液中对其与其他镧系元素混合物中的该元素进行化学计量和伏安法定量的方法报道。这项工作旨在改进一种在乙腈中检测钐的方法,其预期功能是通过差分脉冲伏安法获得钐的选择性电流响应,利用Box-Behnken实验设计来确定测定的最佳条件。具体而言,实验中选择的三个因素,即电位范围、幅度调制和沉积时间,连续发现其最佳条件分别为-1.5至+1.0 V、0.075 V和60 s。观察到最佳条件导致钐的选择性电流响应,检测限为2.25 mg/L,定量限为7.50 mg/L。此外,准确度为98.70%,相对标准偏差(RSD)下的精密度为1.91%,回收率为98.70%。