Gugliandolo Giovanni, Tabandeh Shahin, Rosso Lucia, Smorgon Denis, Fernicola Vito
Department of MIFT, University of Messina, 98166 Messina, Italy.
VTT Technical Research Centre of Finland Ltd., National Metrology Institute VTT MIKES, P.O. Box 1000, FI-02044 VTT, 02150 Espoo, Finland.
Sensors (Basel). 2021 Apr 17;21(8):2844. doi: 10.3390/s21082844.
In this work, the authors exploited the whispering gallery mode (WGM) resonator properties as a thermometer. The sensor is made of a cylindrical sapphire microwave resonator in the center of a gold-plated copper cavity. Two coaxial cables act as antennas and excite the WGM standing waves in the cylindrical sapphire at selected resonance frequencies in the microwave range. The system affords a high quality factor that enables temperature measurements with a resolution better than 15 µK and a measurement standard uncertainty of 1.2 mK, a value approximately three times better than that achieved in previous works. The developed sensor could be a promising alternative to platinum resistance thermometers, both as a transfer standard in industrial applications and as an interpolating instrument for the dissemination of the kelvin.
在这项工作中,作者利用回音壁模式(WGM)谐振器特性作为温度计。该传感器由位于镀金铜腔中心的圆柱形蓝宝石微波谐振器制成。两根同轴电缆充当天线,并在微波范围内的选定共振频率下激发圆柱形蓝宝石中的WGM驻波。该系统具有高品质因数,能够进行分辨率优于15 μK的温度测量,测量标准不确定度为1.2 mK,该值比以前的工作中实现的值大约好三倍。所开发的传感器有望成为铂电阻温度计的替代方案,既可以作为工业应用中的传递标准,也可以作为开尔文温度传递的内插仪器。