Wang Mingwu, Berdahl John, McCafferty Sean
Ophthalmology, Cornea Associates, Tucson, Arizona, USA.
Ophthalmology, Vance Thompson Vision, Sioux Falls, South Dakota, USA.
Curr Eye Res. 2021 Nov;46(11):1614-1620. doi: 10.1080/02713683.2021.1916039. Epub 2021 May 26.
: Evaluate potential corneal biomechanical changes following corneal crosslinking (CXL) by paired differential tonometry intraocular pressure (IOP) measurements with a Goldmann tonometer (GAT) prism and corneal compensating, correcting applanation tonometry surface (CATS) prism.: IOP was measured prospectively on 23 unique eyes undergoing CXL for keratoconus with a GAT using a standard flat GAT prism and a curved corneal error correcting CATS prism before treatment and at 2 weeks, 2 months and 6 months after treatment. Concurrent measurements of central corneal thickness (CCT) and corneal hysteresis (CH) were completed.: Paired IOP measurements with standard GAT and corneal correcting CATS prisms indicated a significant sustained relative increase in the differential IOP between the two prisms after CXL ( = .002,0.051,0.062). CH initially decreased at two weeks post-CXL then returned to sustained pre-op levels ( = .033,0.20,0.20). CCT progressively decreased following CXL ( = .005).: Differential tonometry between standard GAT and corneal biomechanical compensating CATS prisms indicates findings consistent with increased corneal rigidity following CXL and may demonstrate a simple and sensitive method for measurement of relative corneal biomechanical changes due to pharmacologic agents and procedures.
通过使用戈德曼眼压计(GAT)棱镜和角膜补偿校正压平眼压测量表面(CATS)棱镜进行配对差压眼压测量,评估角膜交联(CXL)后潜在的角膜生物力学变化。
前瞻性地对23只接受圆锥角膜CXL治疗的单眼,在治疗前以及治疗后2周、2个月和6个月,使用标准扁平GAT棱镜和弯曲角膜误差校正CATS棱镜,通过GAT测量眼压。同时完成中央角膜厚度(CCT)和角膜滞后(CH)的测量。
使用标准GAT和角膜校正CATS棱镜进行配对眼压测量表明,CXL后两个棱镜之间的差压眼压有显著持续的相对增加(=0.002、0.051、0.062)。CH在CXL后两周时最初下降,然后恢复到术前持续水平(=0.033、0.20、0.20)。CXL后CCT逐渐下降(=0.005)。
标准GAT和角膜生物力学补偿CATS棱镜之间的差压眼压测量表明,其结果与CXL后角膜硬度增加一致,并且可能展示了一种用于测量由于药物和手术导致的相对角膜生物力学变化的简单且敏感的方法。