CALIPER Program, Pediatric Laboratory Medicine, The Hospital for Sick Children, Toronto, Canada.
Laboratory Medicine and Pathobiology, University of Toronto, Toronto, Canada.
Am J Clin Pathol. 2021 Nov 8;156(6):1030-1037. doi: 10.1093/ajcp/aqab064.
Point-of-care testing (POCT) is being increasingly adopted to support clinical care. Data for critical care parameters in healthy children on POCT instruments are lacking. We established comprehensive reference standards for several whole blood parameters on the Radiometer ABL90 FLEX PLUS blood gas analyzer in the Canadian Laboratory Initiative on Paediatric Reference Intervals (CALIPER) cohort.
Approximately 300 healthy children and adolescents (age range, birth to <19 years; sex, boys and girls) were recruited with informed consent. Venous whole blood was collected (using heparinized syringes) and rapidly analyzed at the point of collection for pH, Pco2, Po2, carboxyhemoglobin, methemoglobin, lactate, and electrolytes on the ABL90 FLEX PLUS instrument. Reference intervals were established according to Clinical and Laboratory Standards Institute guidelines.
Of the parameters assessed, 6 required age partitioning; none required sex partitioning. Reference value distributions were consistent across the pediatric age range, demonstrating higher variation in the early neonatal period.
This study established reference standards for 10 critical care analytes in the CALIPER cohort for the first time. These data contribute to our understanding of normative pediatric values for venous electrolytes, metabolites, and blood gases on a modern POCT instrument, facilitating test interpretation in clinical settings that use these assays.
床边检测(POCT)越来越多地被用于支持临床护理。关于 POCT 仪器在健康儿童的关键参数的数据仍然缺乏。我们在加拿大儿科参考区间实验室倡议(CALIPER)队列中建立了 Radiometer ABL90 FLEX PLUS 血气分析仪上多个全血参数的综合参考标准。
使用肝素化注射器采集约 300 名健康儿童和青少年(年龄范围为出生至 <19 岁;性别为男童和女童)的静脉全血,并在采集点快速分析 ABL90 FLEX PLUS 仪器上的 pH、Pco2、Po2、碳氧血红蛋白、高铁血红蛋白、乳酸和电解质。参考区间是根据临床和实验室标准协会的指南确定的。
在所评估的参数中,有 6 个需要按年龄分段;没有需要按性别分段的参数。参考值分布在整个儿科年龄范围内是一致的,在新生儿早期表现出更高的变异。
本研究首次在 CALIPER 队列中为 10 个关键护理分析物建立了参考标准。这些数据有助于我们了解现代 POCT 仪器上静脉电解质、代谢物和血气的正常儿科值,为在使用这些检测的临床环境中进行检测解释提供便利。