Scand J Clin Lab Invest. 2021 Sep;81(5):418-421. doi: 10.1080/00365513.2021.1930142. Epub 2021 May 31.
GEM Premier ChemSTAT is a whole-blood analyzer designed for providing a rapid basic metabolic panel, inclusive of creatinine and blood urea nitrogen, with the unique characteristic of providing measured bicarbonate (HCO) levels. The aim of this work was to evaluate the clinical performance of HCO assessment with this analyser in a real-life hemodialysis setting. Imprecision was calculated at different HCO levels, along with assay comparison with Gem Premier 4000 analysers. GEM Premier ChemSTAT displayed an imprecision and a bias (in comparison to GEM Premier 4000) for HCO of 0.4% and 37.3% at 20.8 mmol/L, 1.2% and 25.6% at 16.4 mmol/L, and 2.1% and 11.6% at 11.5 mmol/L, respectively, using three levels of HCO quality control sample ChemSTAT System Evaluator. At direct comparison with the GEM Premier 4000 in the hemodialysis setting, Bland-Altman analysis of HCO levels evidenced a bias () of -4.9 (95% CI, -5.2 to -4.7) mmol/L. Such difference was attenuated by recalculating the GEM ChemSTAT expected HCO values from pH and pCO using the Henderson Hasselbach equation, =-0.07 (95%CI, -0.19 to 0.05) mmol/L ( = .24). In conclusion, our results show a remarkable difference between the HCO values reported by GEM ChemSTAT or GEM 4000. New reference values for GEM ChemSTAT HCO shall hence be defined according to our findings. We suggest that further investigation and a re-evaluation of the reference range should be made before extending the clinical use of this device.
GEM Premier ChemSTAT 是一款全血分析仪,旨在提供快速的基本代谢面板,包括肌酐和血尿素氮,并具有提供测量碳酸氢盐 (HCO) 水平的独特特点。本研究旨在评估该分析仪在实际血液透析环境中评估 HCO 的临床性能。在不同 HCO 水平下计算精密度,并与 Gem Premier 4000 分析仪进行测定比较。GEM Premier ChemSTAT 在 20.8mmol/L 时显示 HCO 的精密度和偏差(与 Gem Premier 4000 相比)分别为 0.4%和 37.3%,在 16.4mmol/L 时分别为 1.2%和 25.6%,在 11.5mmol/L 时分别为 2.1%和 11.6%,使用三个水平的 HCO 质量控制样本 ChemSTAT 系统评估器。在直接与血液透析环境中的 GEM Premier 4000 进行比较时,HCO 水平的 Bland-Altman 分析显示偏倚 () 为-4.9(95%CI,-5.2 至-4.7)mmol/L。通过使用 Henderson Hasselbach 方程从 pH 和 pCO 重新计算 GEM ChemSTAT 预期的 HCO 值,可以减轻这种差异,= -0.07(95%CI,-0.19 至 0.05)mmol/L( = 0.24)。总之,我们的结果显示 GEM ChemSTAT 或 GEM 4000 报告的 HCO 值之间存在显著差异。根据我们的发现,应定义新的 GEM ChemSTAT HCO 参考值。我们建议在扩展该设备的临床应用之前,应进行进一步的研究和重新评估参考范围。