Abstract

Aimed at preventing heat strain, health problems, and absenteeism among workers with physically demanding occupations, a continuous, accurate, non-invasive measuring system may help such workers monitor their body (core) temperature. The aim of this study is to evaluate the accuracy and explore the usability of the wearable non-invasive Cosinuss° °Temp thermometer. Ear canal temperature was monitored in 49 workers in real-life working conditions. After individual correction, the results of the laboratory and field study revealed high correlations compared to ear canal infrared thermometry for hospital use. After performance of the real-life working tasks, this correlation was found to be moderate. It was also observed that the ambient environmental outdoor conditions and personal protective clothing influenced the accuracy and resulted in unrealistic ear canal temperature outliers. It was found that the Cosinuss° °Temp thermometer did not result in significant interference during work. Therefore, it was concluded that, without a correction factor, the Cosinuss° °Temp thermometer is inaccurate. Nevertheless, with a correction factor, the reliability of this wearable ear canal thermometer was confirmed at rest, but not in outdoor working conditions or while wearing a helmet or hearing protection equipment.

摘要

针对体力劳动职业中工人的热应激、健康问题和旷工问题,连续、准确、非侵入式测量系统可能有助于此类工人监测他们的身体(核心)温度。本研究旨在评估可穿戴式非侵入式 Cosinuss° °Temp 温度计的准确性并探索其可用性。在真实工作条件下监测了 49 名工人的耳道温度。经过个体校正后,实验室和现场研究的结果与医院使用的耳道红外测温仪相比显示出高度相关性。在完成实际工作任务后,发现这种相关性是中等的。还观察到环境户外条件和个人防护装备会影响准确性,并导致不切实际的耳道温度异常值。研究发现,Cosinuss° °Temp 温度计在工作过程中不会造成明显干扰。因此,研究得出结论,没有校正因子,Cosinuss° °Temp 温度计是不准确的。然而,有了校正因子,这种可穿戴式耳道温度计在休息时的可靠性得到了确认,但在户外工作条件下或戴头盔或使用听力保护设备时则不行。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/88b7/8125248/523a512745f0/ijerph-18-04896-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验