Reduction of radiation dose and imaging costs in scoliosis radiography. Application of large-screen image intensifier photofluorography.

作者信息

Manninen H, Kiekara O, Soimakallio S, Vainio J

机构信息

Department of Diagnostic Radiology, University Central Hospital of Kuopio, Finland.

出版信息

Spine (Phila Pa 1976). 1988 Apr;13(4):409-12. doi: 10.1097/00007632-198804000-00007.

Abstract

Photofluorography using a large-field image intensifier (Siemens Optilux 57) was applied to scoliosis radiography and compared with a full-size rare-earth screen/film technique. When scoliosis radiography (PA-projection) was performed on 25 adolescent patients, the photofluorographs were found to be of comparable diagnostic quality with full-size films. A close correspondence between the imaging techniques was found in the Cobb angle measurements as well as in the grading of rotation with the pedicle method. The use of photofluorography results in a radiation dose reduction of about one-half and considerable savings in direct imaging costs and archive space. In our opinion the method is particularly well-suited for follow-up and screening evaluation of scoliosis, but in tall patients the image field size of 40 x 40 cm restricts its usefulness as initial examination.

摘要

使用大型场图像增强器(西门子Optilux 57)的荧光摄影术应用于脊柱侧弯X线摄影,并与全尺寸稀土增感屏/胶片技术进行比较。对25例青少年患者进行脊柱侧弯X线摄影(前后位投照)时,发现荧光照片与全尺寸胶片具有相当的诊断质量。在Cobb角测量以及用椎弓根法进行的旋转分级方面,发现两种成像技术之间有密切的对应关系。使用荧光摄影术可使辐射剂量降低约一半,并在直接成像成本和存档空间方面节省可观。我们认为该方法特别适合脊柱侧弯的随访和筛查评估,但对于身材高大的患者,40×40 cm的图像视野大小限制了其作为初次检查的实用性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验