圆锥角膜患者自动主观验光与传统主观验光的比较

A Comparison between Automated Subjective Refraction and Traditional Subjective Refraction in Keratoconus Patients.

作者信息

Carracedo Gonzalo, Carpena-Torres Carlos, Pastrana Cristina, Privado-Aroco Ana, Serramito María, Espinosa-Vidal Teresa María, Rodríguez-Lafora María

机构信息

Ocupharm Research Group, Department of Optometry and Vision, Faculty of Optics and Optometry, Complutense University of Madrid, Madrid, Spain.

出版信息

Optom Vis Sci. 2021 Jun 1;98(6):597-604. doi: 10.1097/OPX.0000000000001710.

Abstract

SIGNIFICANCE

The performance of the Eye Refract (Luneau Technology, Chartres, France), a new instrument to perform aberrometry-based automated subjective refraction, has been previously evaluated in healthy subjects. However, its clinical implications in other ocular conditions are still unknown.

PURPOSE

The purpose of this study was to evaluate the agreement between the Eye Refract and the traditional subjective refraction, as the criterion standard, in keratoconus patients with and without intracorneal ring segments (ICRSs).

METHODS

A total of 50 eyes of 50 keratoconus patients were evaluated, dividing the sample into 2 groups: 27 eyes without ICRS (37.78 ± 9.35 years) and 23 eyes with ICRS (39.26 ± 13.62 years). An optometrist conducted the refraction with the Eye Refract, and another different optometrist conducted the traditional subjective refraction on the same day. Spherical equivalent (M), cylindrical vectors (J0 and J45), and corrected distance visual acuity were compared between both methods of refraction. In addition, Bland-Altman analysis was performed to assess the agreement between both methods of refraction.

RESULTS

There were no statistically significant differences (P ≥ .05) between the Eye Refract and the traditional subjective refraction for all the variables under study in either group. Without ICRS, the mean difference and 95% limits of agreement (upper, lower) were -0.20 (+1.50, -1.89) D for M, -0.14 (+1.40, -1.68) D for J0, and +0.05 (+1.23, -1.14) D for J45. With ICRS, these values worsened to -0.62 (+3.89, -5.12) D for M, +0.06 (+2.46, -2.34) D for J0, and -0.02 (+2.23, -2.28) D for J45.

CONCLUSIONS

The Eye Refract seems to offer similar results compared with the traditional subjective refraction in keratoconus patients not implanted with ICRS. However, some patients could show abnormal measurements, especially those with ICRS, who should be treated with caution in clinical practice.

摘要

意义

Eye Refract(法国沙特尔的Luneau Technology公司生产)是一种基于像差测量的新型自动主观验光仪,此前已在健康受试者中进行了评估。然而,其在其他眼部疾病中的临床意义仍不明确。

目的

本研究的目的是评估Eye Refract与作为标准参照的传统主观验光在植入和未植入角膜内环(ICRS)的圆锥角膜患者中的一致性。

方法

对50例圆锥角膜患者的50只眼进行评估,将样本分为两组:27只未植入ICRS的眼(年龄37.78±9.35岁)和23只植入ICRS的眼(年龄39.26±13.62岁)。一名验光师使用Eye Refract进行验光,另一名不同的验光师在同一天进行传统主观验光。比较两种验光方法的等效球镜度(M)、柱镜矢量(J0和J45)以及矫正远视力。此外,进行Bland-Altman分析以评估两种验光方法之间的一致性。

结果

在两组研究的所有变量中,Eye Refract与传统主观验光之间均无统计学显著差异(P≥0.05)。未植入ICRS时,M的平均差异和95%一致性界限(上限、下限)为-0.20(+1.50,-1.89)D,J0为-0.14(+1.40,-1.68)D,J45为+0.05(+1.23,-1.14)D。植入ICRS时,这些值分别恶化至M为-0.62(+3.89,-5.12)D,J0为+0.06(+2.46,-2.34)D,J45为-0.02(+2.23,-2.28)D。

结论

在未植入ICRS的圆锥角膜患者中,Eye Refract与传统主观验光相比似乎能提供相似的结果。然而,一些患者可能会出现异常测量结果,尤其是植入ICRS的患者,在临床实践中应谨慎对待。