Repeatability of Aberrometry-Based Automated Subjective Refraction in Healthy and Keratoconus Subjects.

作者信息

Carracedo Gonzalo, Carpena-Torres Carlos, Pastrana Cristina, Privado-Aroco Ana, Serramito María, Batres Laura

机构信息

Department of Optometry and Vision, Faculty of Optics and Optometry, Complutense University of Madrid, Madrid, Spain.

出版信息

J Ophthalmol. 2020 Oct 30;2020:4831298. doi: 10.1155/2020/4831298. eCollection 2020.

Abstract

PURPOSE

To compare the intersession repeatability of the Eye Refract, a new instrument to perform aberrometry-based automated subjective refraction, on healthy and keratoconus subjects.

MATERIALS AND METHODS

A cross-sectional and randomized study was performed. A total of 64 participants were evaluated in the study, selecting one eye per participant randomly. The sample was divided into two different groups: 33 healthy subjects (38.85 ± 13.21 years) and 31 with keratoconus (37.29 ± 11.37 years). Three refractions per participant with the Eye Refract were performed on three different days, without cycloplegia. The repeatability analysis of refractive variables (M, J0, and J45), binocular corrected distance visual acuity (BCDVA), and spent time in refraction was performed in terms of repeatability ( ), its 95% confidence interval (), and intraclass correlation coefficient (ICC).

RESULTS

There were no statistically significant differences ( ≥ 0.05) between sessions in both groups for all refractive variables (M, J0, and J45) and BCDVA. Spent time in refraction was reduced as the sessions went by ( < 0.05). The Eye Refract was more repeatable for refractive errors assessment in healthy subjects (M :   = 0.27 D; J0 :   = 0.09 D; J45 :   = 0.06 D) compared to those with keratoconus (M :   = 0.65 D; J0 :   = 0.29 D; J45 :   = 0.24 D), while it was similar for BCDVA.

CONCLUSIONS

The Eye Refract offered better repeatability to assess refractive errors in healthy subjects compared to those with keratoconus. Despite measurements being also consistent in keratoconus subjects, they should be treated with caution in clinical practice.

摘要

目的

比较新型基于像差的自动主观验光仪Eye Refract在健康受试者和圆锥角膜受试者中的不同测量 session 间的重复性。

材料与方法

进行了一项横断面随机研究。共有64名参与者纳入研究,每名参与者随机选择一只眼睛。样本分为两个不同组:33名健康受试者(38.85±13.21岁)和31名圆锥角膜患者(37.29±11.37岁)。在未使用睫状肌麻痹剂的情况下,每名参与者使用Eye Refract在三个不同日期进行三次验光。对屈光变量(M、J0和J45)、双眼矫正远视力(BCDVA)和验光用时进行重复性分析,分析指标包括重复性标准差( )、其95%置信区间( )和组内相关系数(ICC)。

结果

两组所有屈光变量(M、J0和J45)和BCDVA在各测量 session 间均无统计学显著差异( ≥ 0.05)。随着测量 session 的增加,验光用时减少( < 0.05)。与圆锥角膜患者相比,Eye Refract在健康受试者中对屈光不正评估的重复性更好(M: = 0.27 D;J0: = 0.09 D;J45: = 0.06 D),而在BCDVA方面两者相似(圆锥角膜患者M: = 0.65 D;J0: = 0.29 D;J45: = 0.24 D)。

结论

与圆锥角膜患者相比,Eye Refract在健康受试者中评估屈光不正时具有更好的重复性。尽管在圆锥角膜患者中测量结果也具有一致性,但在临床实践中仍应谨慎对待。