Boijsen M, Goss F, Jacobsson L, Lamm C, Schlossman D, Björneld L, Tylén U
Department of Diagnostic Radiology, Sahlgrenska Sjukhuset, University of Gothenburg, Sweden.
Acta Radiol. 1988 Jul-Aug;29(4):419-22.
Cardiac output was determined with an indicator dilution technique during digital venous angiography of the left ventricle in 11 patients. The contrast medium injected into the right atrium was used as indicator. During and after the injection of contrast medium one blood sample per second was obtained through a catheter placed in the descending aorta. The samples were analyzed for iodine content with x-ray fluorescence analysis and cardiac output determined ad modum Stewart-Hamilton. Thermodilution was used as a reference method. The results indicate that the indicator dilution method with the use of contrast medium might be used for calibration of videodensitometric methods for blood flow measurements.