Higashida Y, Moribe N, Hirata Y, Morita K, Doudanuki S, Sonoda Y, Katsuda N, Hiai Y, Misumi W, Matsumoto M
Department of Radiological Technology, College of Medical Science, Kumamoto University, Japan.
Comput Med Imaging Graph. 1988 May-Jun;12(3):137-45. doi: 10.1016/0895-6111(88)90025-0.
Threshold contrasts of low-contrast objects with computed radiography (CR) images were compared with those of blue and green emitting screen-film systems by employing the 18-alternative forced choice (18-AFC) procedure. The dependence of the threshold contrast on the incident X-ray exposure and also the object size was studied. The results indicated that the threshold contrasts of CR system were comparable to those of blue and green screen-film systems and decreased with increasing object size, and increased with decreasing incident X-ray exposure. The increase in threshold contrasts was small when the relative incident exposure decreased from 1 to 1/4, and was large when incident exposure was decreased further.