Department of Pathology, Faculty of Medicine Ramathibodi Hospital, Mahidol University, Bangkok, Thailand.
Int J Lab Hematol. 2021 Dec;43(6):1373-1378. doi: 10.1111/ijlh.13641. Epub 2021 Jul 8.
Although the microscopic manual count is considered the standard method for NRBC enumeration, modern hematology analyzers can perform this task automatically with reliable accuracy and efficiency. This study aims to evaluate the diagnostic performance of the Sysmex XN hematology analyzer and to construct the optimal workflow for accurate and efficient NRBC reporting.
Specimens containing different levels of NRBC were included. Analytical performance was evaluated via method comparison with flow cytometry (FCM) and manual count (MC). Clinical sensitivity was analyzed by ROC analysis using manual count as the standard method.
Correlation study of %NRBC with FCM and MC demonstrated an r value of 0.925 (95% CI 0.905 to 0.942) and 0.990 (95% CI 0.987 to 0.992) with a mean difference of -0.8 (95%CI: -6.7 to +5.0) and +0.50 (95% CI: -6.7 to +7.7), respectively. When the automated NRBC count was equal to zero and >0.07 × 10 /L, the false-negative rate and false-positive rate were 100%, respectively; hence, manual slide review could be omitted. A false-positive rate of 72.7% was noted in specimens containing NRBC count less than 0.07 × 10 /L.
The Sysmex XN can help improve the efficiency of NRBC enumeration owing to its accuracy, rapidity, and automation. However, further studies are required to improve the accuracy of detection in specimens containing a very low level of NRBC.
虽然显微镜手动计数被认为是 NRBC 计数的标准方法,但现代血液学分析仪可以自动完成这项任务,具有可靠的准确性和效率。本研究旨在评估 Sysmex XN 血液学分析仪的诊断性能,并构建用于准确、高效 NRBC 报告的最佳工作流程。
纳入含有不同水平 NRBC 的标本。通过与流式细胞术(FCM)和手动计数(MC)的方法比较来评估分析性能。使用手动计数作为标准方法进行 ROC 分析来评估临床灵敏度。
%NRBC 与 FCM 和 MC 的相关性研究显示 r 值分别为 0.925(95%CI 0.905 至 0.942)和 0.990(95%CI 0.987 至 0.992),平均差异分别为-0.8(95%CI:-6.7 至 +5.0)和+0.50(95%CI:-6.7 至 +7.7)。当自动 NRBC 计数等于零时和>0.07×10/L 时,假阴性率和假阳性率分别为 100%,因此可以省略手动幻灯片审查。在 NRBC 计数小于 0.07×10/L 的标本中,假阳性率为 72.7%。
由于 Sysmex XN 的准确性、快速性和自动化,它可以帮助提高 NRBC 计数的效率。然而,需要进一步的研究来提高在含有极低水平 NRBC 的标本中的检测准确性。