Development of a filtered AXUV diode array for X-pinch soft x-ray spectra in the energy range of 1-10 keV.

作者信息

Ham Seunggi, Ryu Jonghyeon, Park Sungbin, Chung Kyoung-Jae, Ghim Y-C, Woo H J, Hwang Y S

机构信息

Department of Energy Systems Engineering, Seoul National University, Seoul 08826, South Korea.

Department of Nuclear and Quantum Engineering, KAIST, Daejeon 34141, South Korea.

出版信息

Rev Sci Instrum. 2021 May 1;92(5):053509. doi: 10.1063/5.0034638.

Abstract

We develop a filtered absolute extreme ultraviolet (AXUV) diode array to measure the time evolution of the soft x-ray spectrum in the energy range of 1-10 keV. AXUV-HS5, the detector, has a fast rise time of 0.7 ns, a wide energy detection range, and high accessibility. We use Geant4 simulations to design an appropriate filter set for flat-and-sharp virtual channels (VCs), where a filter with no spectral edge removes large tails of the response curves. A Levenberg-Marquardt (LM) method, sensitive to the expected spectral function, is improved to reliably generate a continuous radiation spectrum, by utilizing spectral information from the least-squares (LS) method that reconstructs a discrete spectrum with low spectral resolution directly from the VCs. We test the filtered AXUV diode array on an X-pinch device with a peak current of 140 kA at Seoul National University; the array with ten channels is installed in a vacuum chamber. For a two-wire 40 μm stainless steel X-pinch, x-ray power, radiation temperature, and the reconstructed x-ray spectrum are obtained from the filtered AXUV diode array by the combined LS-LM method.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验